FDA Adverse Event
Injury
Summary report: N
4 FR S/L GROSHONG NXT REPAIR KIT
MDR report key: 1251283
·
Received December 4, 2008
Report
- Report Number
- 3006260740-2008-00223
- Event Type
- Injury
- Date Received
- December 4, 2008
- Report Date
- November 10, 2008
- Manufacturer
- C. R. BARD INC. (BASD)
- Product Code
- LJS
- PMA / PMN Number
- K034020
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- UK
- Reporter Occupation
- OTHER
Narratives
Additional Manufacturer Narrative · 1
THE DEVICE HAS NOT BEEN RETURNED TO THE MANUFACTURER AT THIS TIME FOR EVALUATION. A CHR REVIEW SHOWED NO OTHER SIMILAR PRODUCT COMPLAINT FILE(S) FROM THIS LOT.
Description of Event or Problem · 1
IT WAS REPORTED THAT A REPAIR WAS PERFORMED ON LEAKING PICC THEN ATTACHED TO 5FU PUMP. IT WAS NOTED LATER THAT THE PICC WAS NOT ATTACHED TO THE CATHETER HUB. THE HUB REMAINED ATTACHED TO THE STATLOCK, AND THE PICC CATHETER HAD MIGRATED INTO THE PATIENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | 4 FR S/L GROSHONG NXT REPAIR KIT | LJS | C. R. BARD INC. (BASD) | RESF1134 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Required Intervention |