FDA Adverse Event Death Summary report: N

X SERIES

MDR report key: 12509162 · Received September 21, 2021

Report

Report Number
1220908-2021-03017
Event Type
Death
Date Received
September 21, 2021
Date of Event
August 30, 2021
Report Date
August 31, 2021
Manufacturer
ZOLL MEDICAL CORPORATION
Product Code
MKJ
UDI-DI
00847946004354
PMA / PMN Number
K112432/P160
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
CA, US
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 0

THE DEVICE WAS RETURNED TO ZOLL MEDICAL CORPORATION. THE CUSTOMER'S REPORT WAS DUPLICATED AND ATTRIBUTED TO A MISSING GASKET IN THE MULTIFUNCTION CABLE. THE DEVICE WAS RECERTIFIED AND RETURNED TO THE CUSTOMER. ANALYSIS OF REPORTS OF THIS TYPE HAS NOT IDENTIFIED AN INCREASE IN TREND.

Additional Manufacturer Narrative · 1

ZOLL MEDICAL CORPORATION HAS RECEIVED THE PRODUCT AND WILL BE PROVIDING A SUPPLEMENTAL REPORT WHEN OUR INVESTIGATION IS COMPLETED.

Description of Event or Problem · 1

COMPLAINANT ALLEGED THAT WHILE ATTEMPTING TO DEFIBRILLATE A FEMALE PATIENT (AGE UNKNOWN), THE DEVICE WAS UNABLE TO OBTAIN AN ECG SIGNAL VIA ELECTRODE PADS. COMPLAINANT INDICATED THAT THE PATIENT SUBSEQUENTLY EXPIRED, HOWEVER IT WAS NOT A RESULT OF THE REPORTED MALFUNCTION.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1406596 X SERIES DEFIBRILLATOR/PACEMAKER MKJ ZOLL MEDICAL CORPORATION X SERIES NA 00847946004354

Patients

Seq Age Sex Outcome Treatment
1 Death