FDA Adverse Event
Malfunction
Summary report: N
TOTALCARE BARIATRIC
MDR report key: 1250897
·
Received December 2, 2008
Report
- Report Number
- 1824206-2008-04449
- Event Type
- Malfunction
- Date Received
- December 2, 2008
- Date of Event
- October 24, 2008
- Report Date
- October 24, 2008
- Manufacturer
- HILL-ROM
- Product Code
- FNL
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- NM, US
Narratives
Additional Manufacturer Narrative · 1
THE TECHNICIAN INVESTIGATED AND FOUND THE FOOT HI/LOW UP SOLENOID PLUNGER WAS STUCK. THE TECHNICIAN COULD NOT REPAIR THE BED DUE TO THE BED BEING IN USE. THE FACILITIES STAFF IS AWARE THE BED WILL NOT GO INTO TRENDELENBURG AND DECIDED NOT TO MOVE THE PATIENT.
Description of Event or Problem · 1
INFORMATION RECEIVED INDICATES THE TRENDELENBURG FUNCTION WILL NOT WORK.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | TOTALCARE BARIATRIC | AC POWERED HOSPITAL BED | FNL | HILL-ROM | 1840 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | UNK |