FDA Adverse Event Malfunction Summary report: N

PUMP MMT-515NAS PRDGM INS V2.1 SK EN

MDR report key: 1250866 · Received November 26, 2008

Report

Report Number
3004209178-2008-00995
Event Type
Malfunction
Date Received
November 26, 2008
Date of Event
November 7, 2008
Report Date
November 7, 2008
Manufacturer
MEDTRONIC PUERTO RICO OPERATIONS MED-REL
Product Code
LZG
PMA / PMN Number
K040676
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CA, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

CURRENTLY, IT IS UNK WHETHER OR NOT THE DEVICE MAY HAVE CAUSED OR CONTRIBUTED TO THE EVENT. THE DEVICE HAS BEEN RETURNED, BUT NOT YET EVALUATED. NO CONCLUSION CAN BE DRAWN AT THIS TIME. FURTHER INFORMATION WILL FOLLOW ONCE THE ANALYSIS HAS BEEN COMPLETED.

Description of Event or Problem · 1

THE CUSTOMER STATED THAT THE INSULIN PUMP ALARMED MOTOR ERROR. TROUBLESHOOTING WAS PERFORMED, AND THE INSULIN PUMP FAILED THE DISPLACEMENT TEST. THE CUSTOMER WAS ADVISED TO REVERT TO A BACKUP PLAN TO TREAT HIS DIABETES. NOTHING FURTHER WAS REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 PUMP MMT-515NAS PRDGM INS V2.1 SK EN INFUSION PUMP LZG MEDTRONIC PUERTO RICO OPERATIONS MED-REL MMT-515NAS

Patients

Seq Age Sex Outcome Treatment
1 66 YR