FDA Adverse Event
Malfunction
Summary report: N
THERMOPHORE ARTHRITIS PAD
MDR report key: 1250727
·
Received December 1, 2008
Report
- Report Number
- 1811605-2008-00147
- Event Type
- Malfunction
- Date Received
- December 1, 2008
- Date of Event
- October 1, 2008
- Report Date
- November 14, 2008
- Manufacturer
- BATTLE CREEK EQUIPMENT COMPANY
- Product Code
- IRT
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- MI, US
- Reporter Occupation
- NOT APPLICABLE
Narratives
Description of Event or Problem · 1
IT WAS REPORTED THE FABRIC FOUNDATION OF THE UNIT WAS BURNED. INSPECTION OF THE UNIT REVEALED SIGNIFICANT DAMAGE TO THE FABRIC FOUNDATION AND COVER OF THE HEATING UNIT CAUSED BY EXCESSIVE LEVELS OF HEAT, HOWEVER, THERE WAS NO EVIDENCE OF AN ACTUAL BURN. THE CONDITION OF THE UNIT SUGGESTED THAT THE UNIT HAD BEEN FOLDED DURING USE, COMPRESSING AND BREAKING THE COMPONENT. WE CONCLUDED THAT THIS REPORT WAS ATTRIBUTABLE TO MISUSE ON THE PART OF THE CONSUMER.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | THERMOPHORE ARTHRITIS PAD | ELECTRIC MOIST HEAT PACK | IRT | BATTLE CREEK EQUIPMENT COMPANY | 156 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |