FDA Adverse Event Malfunction Summary report: N

TOTALCARE

MDR report key: 1250641 · Received November 14, 2008

Report

Report Number
1824206-2008-04192
Event Type
Malfunction
Date Received
November 14, 2008
Date of Event
October 14, 2008
Report Date
October 14, 2008
Manufacturer
HILL-ROM RITTER
Product Code
FNL
Product Problem
Yes
Report Source
Manufacturer report

Narratives

Additional Manufacturer Narrative · 1

THE HILL-ROM TECH FOUND THE SCALE WAS OUT OF CALIBRATION. THE TECH CALIBRATED THE SCALE TO REPAIR THE BED.

Description of Event or Problem · 1

INFO RECEIVED INDICATES THE SCALE IS NOT ACCURATE, WILL NOT ZERO AND THE PT'S WEIGHT DISAPPEARS FROM THE SCREEN AFTER THE PT IS WEIGHED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 TOTALCARE AC POWERED HOSPITAL BED FNL HILL-ROM RITTER 1900

Patients

Seq Age Sex Outcome Treatment
1 UNK