FDA Adverse Event Malfunction Summary report: N

ULTRACISION HARMONIC SCALPEL CURVED SHEARS (LCS) - PISTOL GRIP 36CM-15MM ACTIVE

MDR report key: 1250623 · Received November 14, 2008

Report

Report Number
3005075853-2008-03050
Event Type
Malfunction
Date Received
November 14, 2008
Date of Event
October 14, 2008
Report Date
October 14, 2008
Manufacturer
ETHICON ENDO-SURGERY, LLC
Product Code
GEI
PMA / PMN Number
K002981
Removal / Correction Number
NA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
JA
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

EVALUATION SUMMARY: THE DEVICE WAS RETURNED WITH THE BLADE SCRATCHED AND CRACKED. DURING FUNCTIONAL TESTING, AN ERROR CODE 5 WAS DISPLAYED AND THE BLADE TIP FURTHER CRACKED. THE DEVICE WILL STOP ACTIVATING, AND EITHER EMIT A SOLID TONE OR DISPLAY AN INSTRUMENT ERROR CODE 5 OR A SOLID TONE ON THE GENERATOR DISPLAY WHEN THE BLADE BECOMES DAMAGED. WHEN A BLADE HAS BEEN COMPROMISED SUCH AS SCRATCHED OR CRACKED, THE TITANIUM METAL IS FATIGUED WHEN CONTINUALLY ENERGIZED AND THIS RESULTS IN THE TABLE FURTHER CRACKING AND POSSIBLY BREAKING OFF. A POSSIBLE CAUSE OF AN ERROR CODE 5 (INSTRUMENT ERROR) IS BLADE DAMAGE. BLADE DAMAGE MAY OCCUR FROM EXTERNAL CONTACT WITH METAL OR PLASTIC DURING PRE-OP OR GENERAL USE. EACH DEVICE IS VISUALLY INSPECTED AND FUNCTIONALLY TESTED PRIOR TO SHIPMENT AND DAMAGE OF THIS MAGNITUDE WOULD HAVE BEEN DETECTED DURING THIS INSPECTION AND TESTING. HOWEVER, A CORRECTIVE AND PREVENTIVE ACTION HAS BEEN INITIATED TO ADDRESS THIS ISSUE. COMPLAINT INFO IS TRENDED ON A REGULAR BASIS TO DETERMINE IF FURTHER INVESTIGATION IS WARRANTED. THE BATCH HISTORY RECORDS WERE REVIEWED WITH NO ANOMALIES NOTED DURING THE MFG PROCESS.

Description of Event or Problem · 1

IT WAS REPORTED THAT DURING A LOBECTOMY PROCEDURE, THE DEVICE LOCKED OUT. ANOTHER DEVICE WAS USED TO COMPLETE THE CASE. THERE WERE NO ADVERSE CONSEQUENCES TO THE PT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 ULTRACISION HARMONIC SCALPEL CURVED SHEARS (LCS) - PISTOL GRIP 36CM-15MM ACTIVE GEI ETHICON ENDO-SURGERY, LLC NA UNK

Patients

Seq Age Sex Outcome Treatment
1 GENERATOR| HANDPIECE