FDA Adverse Event Malfunction Summary report: N

HARMONIC ACE 36CM W ERG HANDLE

MDR report key: 1250618 · Received November 14, 2008

Report

Report Number
3005075853-2008-03054
Event Type
Malfunction
Date Received
November 14, 2008
Date of Event
October 17, 2008
Report Date
October 17, 2008
Manufacturer
ETHICON ENDO-SURGERY, LLC
Product Code
LFL
PMA / PMN Number
K042777
Removal / Correction Number
NA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
NJ, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

EVALUATION SUMMARY: THE DEVICE WAS RETURNED WITH THE BLADE SCRATCHED AND CRACKED. DURING FUNCTIONAL TESTING, AN ERROR CODE 5 WAS DISPLAYED AND THE BLADE TIP BROKE OFF. THE DEVICE WILL STOP ACTIVATING, AND EITHER EMIT A SOLID TONE OR DISPLAY AN INSTRUMENT ERROR CODE 5 OR A SOLID TONE ON THE GENERATOR DISPLAY WHEN THE BLADE BECOMES DAMAGED. WHEN A BLADE HAS BEEN COMPROMISED, SUCH AS SCRATCHED OR CRACKED, THE TITANIUM METAL IS FATIGUED WHEN CONTINUALLY ENERGIZED AND THIS RESULTS IN THE BLADE FURTHER CRACKING AND POSSIBLY BREAKING OFF. A POSSIBLE CAUSE OF AN ERROR CODE 5 (INSTRUMENT ERROR) IS BLADE DAMAGE. BLADE DAMAGE MAY OCCUR FROM EXTERNAL CONTACT WITH METAL OR PLASTIC DURING PRE-OP OR GENERAL USE. EACH DEVICE IS VISUALLY INSPECTED AND FUNCTIONALLY TESTED PRIOR TO SHIPMENT, AND DAMAGE OF THIS MAGNITUDE WOULD HAVE BEEN DETECTED DURING THIS INSPECTION AND TESTING. HOWEVER, A CORRECTIVE AND PREVENTIVE ACTION HAS BEEN INITIATED TO ADDRESS THIS ISSUE. COMPLAINT INFORMATION IS TRENDED ON A REGULAR BASIS TO DETERMINE IF FURTHER INVESTIGATION IS WARRANTED. THE BATCH HISTORY RECORDS WERE REVIEWED WITH NO ANOMALIES NOTED DURING THE MANUFACTURING PROCESS.

Description of Event or Problem · 1

IT WAS REPORTED THAT DURING A LSH PROCEDURE, THE DEVICE KEPT GIVING A SOLID TONE. THE SYSTEM WENT INTO STANDBY. THERE WAS NO ERROR CODE GIVEN. A NEW DEVICE WAS USED TO COMPLETE THE PROCEDURE. THERE WAS NOT PT CONSEQUENCE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 HARMONIC ACE 36CM W ERG HANDLE LFL ETHICON ENDO-SURGERY, LLC NA E4LM6C

Patients

Seq Age Sex Outcome Treatment
1 HANDPIECE| GENERATOR