HARMONIC ACE 36 CM
Report
- Report Number
- 3005075853-2008-03041
- Event Type
- Malfunction
- Date Received
- November 14, 2008
- Date of Event
- October 16, 2008
- Report Date
- October 21, 2008
- Manufacturer
- ETHICON ENDO-SURGERY, LLC
- Product Code
- LFL
- PMA / PMN Number
- K042777
- Removal / Correction Number
- NA
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- JA
- Reporter Occupation
- OTHER
Narratives
EVALUATION SUMMARY: THE DEVICE WAS TESTED FOR HAND-ACTIVATION AND THE HAND CONTROL SWITCH ASSEMBLY BUTTONS WERE NON-FUNCTIONAL. HOWEVER, WHEN THE DEVICE WAS TESTED USING THE FOOTSWITCH, THE DEVICE ACTIVATED EACH DEVICE IS VISUALLY INSPECTED AND FUNCTIONALLY TESTED PRIOR TO SHIPMENT AND DAMAGE OF THIS MAGNITUDE WOULD HAVE BEEN DETECTED DURING THIS INSPECTION AND TESTING. HOWEVER, A CORRECTIVE AND PREVENTIVE ACTION HAS BEEN INITIATED TO ADDRESS THIS ISSUE. COMPLAINT INFORMATION IS TRENDED ON A REGULAR BASIS TO DETERMINE IF FURTHER INVESTIGATION IS WARRANTED. THE BATCH HISTORY RECORDS WERE REVIEWED WITH NO ANOMALIES NOTED DURING THE MANUFACTURING PROCESS.
IT WAS REPORTED THAT PRIOR TO AN UNKNOWN PROCEDURE, THE HAND SWITCH WAS NON-FUNCTIONAL WHEN THE DEVICE WAS CHECKED. ANOTHER INSTRUMENT WAS USED TO COMPLETE THE PROCEDURE WITH NO PATIENT CONSEQUENCE.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | HARMONIC ACE 36 CM | LFL | ETHICON ENDO-SURGERY, LLC | NA | D4JT44 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | GENERATOR| HANDPIECE |