LIFEVEST WCD 3100 SYSTEM
MDR report key: 1250528
·
Received November 14, 2008
Report
- Report Number
- 3002158293-2008-00604
- Event Type
- Malfunction
- Date Received
- November 14, 2008
- Date of Event
- October 13, 2008
- Report Date
- October 23, 2008
- Manufacturer
- ZOLL LIFECOR CORPORATION
- Product Code
- MVK
- PMA / PMN Number
- P010030
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- TX, US
Narratives
Additional Manufacturer Narrative · 1
DEVICE EVALUATION SUMMARY: DEVICE EVALUATION OF MONITOR HAS BEEN COMPLETED. THE REPORTED PROBLEM WAS CONFIRMED. THE MONITOR WAS FOUND TO HAVE A DEFECTIVE BATTERY CONNECTOR. THE CONNECTOR PINS WERE BENT. THE CONNECTOR WAS REPAIRED. THE MONITOR WAS RETESTED AND THEN RESTOCKED. THE ROOT CAUSE OF THE BENT PINS IS UNK, BUT IS LIKELY DUE TO A BATTERY PACK BEING FORCED INTO A MONITOR WITH THE CONNECTORS MISALIGNED. THE DAMAGED CONNECTOR ON THE MONITOR WAS REPLACED. NO ADVERSE EVENTS RESULTED FROM THE DAMAGED MONITOR. THE PSR RECEIVED A REPLACEMENT MONITOR.
Description of Event or Problem · 1
A LIFECOR PT SERVICES REP (PSR) CONTACTED LIFECOR CUSTOMER SUPPORT TO REPORT THAT THE PINS IN HER MONITOR WERE BENT. SUPPORT SENT THE PSR A REPLACEMENT MONITOR.