FREESTYLE LIBRE 2
Report
- Report Number
- 2954323-2021-85084
- Event Type
- Injury
- Date Received
- September 20, 2021
- Date of Event
- August 29, 2021
- Report Date
- September 20, 2021
- Manufacturer
- ABBOTT DIABETES CARE INC
- Product Code
- QLG
- PMA / PMN Number
- K193371
- Adverse Event
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- SP
- Reporter Occupation
- 003
Narratives
THE REPORTED PRODUCT IS NOT EXPECTED TO BE RETURNED AS THE DEVICE WAS REPORTEDLY DISCARDED. AN EXTENDED INVESTIGATION HAS BEEN PERFORMED FOR THE REPORTED COMPLAINT. THERE WAS NO INDICATION THAT THE PRODUCT DID NOT MEET SPECIFICATION. THE REPORTED COMPLAINT IS RELATED TO SKIN IRRITATION OR AN ALLERGIC REACTION TO THE PATCH ADHESIVE OF THE FREESTYLE LIBRE SENSOR. DHRS (DEVICE HISTORY RECORD) FOR THE FREESTYLE LIBRE SENSOR AND FREESTYLE LIBRE SENSOR KIT WERE REVIEWED, AND THE DHRS SHOWED THE FREESTYLE LIBRE SENSOR AND FREESTYLE LIBRE SENSOR KIT PASSED ALL TESTS PRIOR TO RELEASE. DOSE AUDIT REPORTS WERE REVIEWED AND DEMONSTRATES THE CONTINUED EFFECTIVENESS OF THE ESTABLISHED STERILIZATION PROCESS FOR LIBRE SENSOR PRODUCTS. ENVIRONMENTAL MONITORING REPORTS WERE REVIEWED, INCLUDING BIOBURDEN AND ENDOTOXIN TESTING, AND DEMONSTRATED THAT ALL MONITORING PROCESSES CONTINUE TO MEET ADC MINIMUM REQUIREMENTS FOR PRODUCT QUALITY. IN THE EVENT THAT UNANTICIPATED PRODUCT IS RECEIVED, A PHYSICAL INVESTIGATION WILL BE PERFORMED PER ADC'S ESTABLISHED PROCESSES AND PROCEDURES AND A FOLLOW-UP REPORT WILL BE SUBMITTED UPON COMPLETION OF INVESTIGATION. ALL PERTINENT INFORMATION AVAILABLE TO ABBOTT DIABETES CARE HAS BEEN SUBMITTED.
A CAREGIVER REPORTED THE CUSTOMER EXPERIENCED A SKIN REACTION AT THE SITE OF THE FREESTYLE LIBRE 2 SENSOR. THE CUSTOMER EXPERIENCED PUS AND PAIN AT SENSOR SITE, AND WAS PRESCRIBED ANTIBIOTIC OINTMENT LIADE 15 G BY A HEALTHCARE PROVIDER FOR TREATMENT. THERE WAS NO REPORT OF DEATH OR PERMANENT INJURY ASSOCIATED WITH THIS EVENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1397828 | FREESTYLE LIBRE 2 | FLASH GLUCOSE MONITORING SYSTEM | QLG | ABBOTT DIABETES CARE INC | 71992-01 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Required Intervention |