BD BACTEC¿ PLUS AEROBIC/F CULTURE VIALS (PLASTIC)
Report
- Report Number
- 2647876-2021-00079
- Event Type
- Malfunction
- Date Received
- September 20, 2021
- Date of Event
- August 24, 2021
- Report Date
- October 22, 2021
- Manufacturer
- BECTON DICKINSON CARIBE LTD.
- Product Code
- MDB
- UDI-DI
- 00382904420239
- PMA / PMN Number
- K113558
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- MD, US
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
- Health Professional
- Yes
Narratives
H6: INVESTIGATION SUMMARY: CUSTOMER REPORTED BLOOD UNDER THE SENSOR. TWO (2) PHOTO WERE RECEIVED. BD WAS UNABLE TO DUPLICATE CUSTOMER¿S EXPERIENCE WITH THE BACTEC PRODUCT. RETENTION SAMPLES WERE VISUALLY INSPECTED WITH SATISFACTORY RESULTS. BATCH/SENSOR HISTORY RECORDS WERE REVIEWED, AND ALL TESTING WERE WITHIN SPECIFICATION FOR PRODUCT RELEASE. BLOOD BACKGROUND WAS PERFORMED WITH SATISFACTORY RESULTS. A COMPLAINT HISTORY REVIEW WAS CONDUCTED AND ONLY THE CURRENT COMPLAINT WAS FOUND RELATING TO THE INCIDENT LOT NUMBER AND THE ¿AS REPORTED¿ DEFECT CODE. SENSOR ADHESION SCRAPE TEST IS PERFORMED TO EACH SENSOR BATCH AS PART OF RELEASE CRITERIA. COMPLAINT IS CONFIRMED BASED ON PHOTOS RECEIVED FOR SENSOR ADHESION DEFECT. THE VISION SYSTEM IS CHALLENGED PRIOR EACH LOT. PRODUCT INSERT WARNINGS AND PRECAUTIONS SECTIONS STATES THAT PRIOR TO USE, EACH VIAL SHOULD BE EXAMINED FOR EVIDENCE OF CONTAMINATION SUCH AS CLOUDINESS, BULGING OR DEPRESSES SEPTUM, OR LEAKAGE. VIALS SHOWING EVIDENCE OF CONTAMINATION SHOULD NOT BE USED. ALSO PRIOR TO USE, THE USER SHOULD EXAMINE THE VIAL FOR EVIDENCE OF DAMAGE OR DETERIORATION. VIALS DISPLAYING TURBIDITY, CONTAMINATION, OR DISCOLORATION (DARKENING) SHOULD NOT BE USED. CORRECTIVE ACTIONS PREVENTIVE ACTIONS (CAPA) NO. 2882676 WAS INITIATED TO FURTHER INVESTIGATE THESE TYPES OF COMPLAINTS AND DETERMINE ANY APPROPRIATE ACTIONS TO REDUCE THEIR OCCURRENCE. A CROSS FUNCTIONAL TEAM CONTINUALLY MONITORS ALL PRODUCT COMPLAINTS FOR TRENDS AND DETERMINES IF ANY ADDITIONAL ACTIONS ARE NECESSARY BEYOND THE CURRENT INVESTIGATIONAL PROCESS. H3 OTHER TEXT : SEE H10.
IT WAS REPORTED THAT 5 BD BACTEC¿ PLUS AEROBIC/F CULTURE VIALS (PLASTIC) BOTTLES HAD BAD SENSORS AND PRODUCED FALSE POSITIVES. THERE WAS NO INDICATION THAT RESULTS WERE REPORTED, AND THERE WAS NO REPORT OF PATIENT IMPACT. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER: "CUSTOMER IS REPORTING 5 BOTTLE WITH BAD SENSORS"
THE REPORTED LOT # 118614 WAS NOT FOUND FOR THE REPORTED CATALOG # 442023. MEDICAL DEVICE EXPIRATION DATE: UNKNOWN. A DEVICE EVALUATION IS ANTICIPATED, BUT HAS NOT YET BEGUN. UPON COMPLETION OF THE INVESTIGATION, A SUPPLEMENTAL REPORT WILL BE FILED. DEVICE MANUFACTURE DATE: UNKNOWN.
IT WAS REPORTED THAT 5 BD BACTEC¿ PLUS AEROBIC/F CULTURE VIALS (PLASTIC) BOTTLES HAD BAD SENSORS AND PRODUCED FALSE POSITIVES. THERE WAS NO INDICATION THAT RESULTS WERE REPORTED, AND THERE WAS NO REPORT OF PATIENT IMPACT. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER: "CUSTOMER IS REPORTING 5 BOTTLE WITH BAD SENSORS."