FDA Adverse Event Malfunction Summary report: N

CD HORIZON SPINAL SYSTEM

MDR report key: 12493994 · Received September 20, 2021

Report

Report Number
1030489-2021-01175
Event Type
Malfunction
Date Received
September 20, 2021
Date of Event
August 26, 2021
Report Date
October 21, 2021
Manufacturer
WARSAW ORTHOPEDICS
Product Code
NKB
PMA / PMN Number
SEE H10
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
JA
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 0

H6: INFORMATION RECEIVED THAT PRODUCT REMAINS IMPLANTED HENCE CODE WAS UPDATED. D6B: EXPLANT DATE IS NOT APPLICABLE SINCE THE PRODUCT REMAINS IN PATIENT. MEDTRONIC IS SUBMITTING THIS REPORT TO COMPLY WITH FDA REPORTING REGULATIONS UNDER 21 CFR PARTS 4 AND 803. THIS REPORT IS BASED UPON INFORMATION OBTAINED BY MEDTRONIC, WHICH THE COMPANY MAY NOT HAVE BEEN ABLE TO FULLY INVESTIGATE OR VERIFY PRIOR TO THE DATE THE REPORT WAS REQUIRED BY THE FDA. MEDTRONIC HAS MADE REASONABLE EFFORTS TO OBTAIN MORE COMPLETE INFORMATION AND HAS PROVIDED AS MUCH RELEVANT INFORMATION AS IS AVAILABLE TO THE COMPANY AS OF THE SUBMISSION DATE OF THIS REPORT. THIS REPORT DOES NOT CONSTITUTE AN ADMISSION OR A CONCLUSION BY FDA, MEDTRONIC, OR ITS EMPLOYEES THAT THE DEVICE, MEDTRONIC, OR ITS EMPLOYEE CAUSED OR CONTRIBUTED TO THE EVENT DESCRIBED IN THE REPORT. IN PARTICULAR, THIS REPORT DOES NOT CONSTITUTE AN ADMISSION BY ANYONE THAT THE PRODUCT DESCRIBED IN THIS REPORT HAS ANY ¿DEFECTS¿ OR HAS ¿MALFUNCTIONED¿. THESE WORDS ARE INCLUDED IN THE FDA 3500A FORM AND ARE FIXED ITEMS FOR SELECTION CREATED BY THE FDA TO CATEGORIZE THE TYPE OF EVENT SOLELY FOR THE PURPOSE OF REGULATORY REPORTING. MEDTRONIC OBJECTS TO THE USE OF THESE WORDS AND OTHERS LIKE THEM BECAUSE OF THE LACK OF DEFINITION AND THE CONNOTATIONS IMPLIED BY THESE TERMS. THIS STATEMENT SHOULD BE INCLUDED WITH ANY INFORMATION OR REPORT DISCLOSED TO THE PUBLIC UNDER THE FREEDOM OF INFORMATION ACT. ANY REQUIRED FIELDS THAT ARE UNPOPULATED ARE BLANK BECAUSE THE INFORMATION IS CURRENTLY UNKNOWN OR UNAVAILABLE. A GOOD FAITH EFFORT WILL BE MADE TO OBTAIN THE APPLICABLE INFORMATION RELEVANT TO THE REPORT. IF INFORMATION IS PROVIDED IN THE FUTURE, A SUPPLEMENTAL REPORT WILL BE ISSUED.

Additional Manufacturer Narrative · 1

THIS DEVICE IS NOT APPROVED FOR SALE IN US BUT A SIMILAR DEVICE WITH CATALOG# 1556200500, 510K# K131321 AND UDI (B)(4) IS APPROVED FOR SALE IN US. THIS EVENT IS SERIOUS INJURY REPORTABLE. SINCE THE DEVICE IS NOT MARKETED IN THE US, THE DEVICE IS REPORTED AS A MALFUNCTION FOR A SIMILAR DEVICE. IF INFORMATION IS PROVIDED IN THE FUTURE, A SUPPLEMENTAL REPORT WILL BE ISSUED.

Description of Event or Problem · 1

INFORMATION WAS RECEIVED FROM THE HEALTHCARE PROVIDER (HCP) VIA THE MANUFACTURER REPRESENTATIVE REGARDING AN EVENT WHICH OCCURRED AFTER AN INITIAL T9 / S2 PSF SURGERY ON (B)(6) 2021, DUE TO SPINAL KYPHOSCOLIOSIS. IT WAS REPORTED THAT THE RODS ON BOTH SIDES WERE BROKEN AT LOWER L5 THREE MONTHS AFTER THE INITIAL OPERATION. THE RODS WERE REPLACED AND REINFORCED. IN ADDITION, THE NUTS OF THE S2AI SCREW WERE LOOSE ON BOTH SIDES, SO THEY WERE REPLACED. THE SCREWS HAVE BEEN DISCARDED. NO PATIENT INJURY WAS REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1397325 CD HORIZON SPINAL SYSTEM THORACOLUMBOSACRAL PEDICLE SCREW SYSTEM NKB WARSAW ORTHOPEDICS 1556300500 0801116W

Patients

Seq Age Sex Outcome Treatment
1 61 YR Required Intervention