INSTINCT PLUS ENDOSCOPIC CLIPPING DEVICE
Report
- Report Number
- 1037905-2021-00512
- Event Type
- Malfunction
- Date Received
- September 16, 2021
- Report Date
- September 16, 2021
- Manufacturer
- COOK ENDOSCOPY
- Product Code
- PKL
- UDI-DI
- 00827002580107
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- FL, US
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
Narratives
510(K): K19269. INVESTIGATION EVALUATION: A PRODUCT EVALUATION WAS NOT PERFORMED IN RESPONSE TO THIS REPORT BECAUSE THE PRODUCT SAID TO BE INVOLVED WAS NOT PROVIDED TO COOK FOR EVALUATION. THE REPORT WAS CONFIRMED BASED SOLELY ON THE CUSTOMER'S PHOTO. THE CUSTOMER PROVIDED TWO PHOTOS. ONE OF THE PICTURES, SHOWS THE PRODUCT POUCH. THE LOT NUMBER NUMBER AND RPN MATCHES THE INFORMATION IN THE REPORT. THE SECOND PHOTO SHOWS THE CLIP ATTACHED TO THE DRIVE WIRE IN THE CLOSED POSITION. THE CLIP HAS TROMBONED AND PART OF THE DISTAL END OF THE DRIVE WIRE IS STICKING OUT THE BACK OF THE CLIP. THE DEVICE HISTORY RECORD FOR THE LOT NUMBER SAID TO BE INVOLVED WAS REVIEWED. A DISCREPANCY OR ANOMALY WAS NOT OBSERVED WITH THE PRODUCT THAT WAS RELEASED FOR DISTRIBUTION. INVESTIGATION CONCLUSION: WE COULD NOT CONDUCT A COMPLETE INVESTIGATION BECAUSE THE PRODUCT SAID TO BE INVOLVED WAS NOT RETURNED FOR EVALUATION. A DEFINITIVE CAUSE FOR THE REPORTED OBSERVATION COULD NOT BE DETERMINED. THE COMPLAINT IS CONFIRMED BASED ON THE CUSTOMER'S PHOTO OF THE CLIP. THE INSTRUCTIONS FOR USE STATES: "UNCOIL DEVICE. VERIFY SMOOTH HANDLE OPERATION AND CLIP ACTION. OPEN CLIP BY GENTLY MOVING HANDLE SPOOL DISTALLY (AWAY FROM HANDLE THUMB RING). ONCE CLIP IS FULLY OPEN, DO NOT CONTINUE ADVANCING HANDLE SPOOL AS CLIP MAY PREMATURELY DETACH FROM CATHETER. CLOSE CLIP BY MOVING HANDLE SPOOL PROXIMALLY UNTIL CLIP IS FULLY CLOSED. PRECAUTION: DO NOT CONTINUE TO PULL HANDLE SPOOL BEYOND TACTILE RESISTANCE AS THIS MAY PREMATURELY DEPLOY CLIP." THE INSTRUCTIONS FOR USE STATES: " WITH CLIP CLOSED AND WITHOUT HOLDING HANDLE SPOOL, ADVANCE DEVICE IN SMALL INCREMENTS INTO ACCESSORY CHANNEL OF GASTROSCOPE, DUODENOSCOPE, OR COLONOSCOPE. NOTE: IF DIFFICULT TO ADVANCE DEVICE, RELAX ELEVATOR AND/OR STRAIGHTEN ENDOSCOPE.¿ THE INSTRUCTIONS FOR USE ALSO INCLUDES THE FOLLOWING PRECAUTION: ¿HOLDING HANDLE SPOOL DURING CLIP ADVANCEMENT MAY PREMATURELY DEPLOY CLIP.¿ FAILURE TO FOLLOW THE INSTRUCTIONS ABOVE CAN RESULT IN DAMAGE TO THE DEVICE WHICH MAY LEAD TO PREMATURE CLIP DEPLOYMENT. PRIOR TO DISTRIBUTION, ALL INSTINCT PLUS ENDOSCOPIC CLIPPING DEVICE ARE SUBJECTED TO A VISUAL INSPECTION AND FUNCTIONAL TESTING TO ENSURE DEVICE INTEGRITY. A REVIEW OF THE DEVICE HISTORY RECORD CONFIRMED THAT THE LOT SAID TO BE INVOLVED MET ALL MANUFACTURING REQUIREMENTS PRIOR TO SHIPMENT. CORRECTIVE ACTION: CORRECTIVE ACTION IS NOT WARRANTED AT THIS TIME BASED ON THE QUALITY ENGINEERING RISK ASSESSMENT. QUALITY ASSURANCE WILL CONTINUE TO MONITOR FOR COMPLAINT TRENDS AND REASSESS THE RISK ASSESSMENT RESULTS AS POST MARKET FEEDBACK CONTINUES TO BECOME AVAILABLE.
DURING AN ENDOSCOPIC CLIPPING PROCEDURE, THE PHYSICIAN USED A COOK INSTINCT PLUS ENDOSCOPIC CLIPPING DEVICE. PER THE CUSTOMER, THE "WIRE IS EXPOSED AND THE CLIP DID NOT OPEN." A PHOTO PROVIDED DEPICTS THE CLIP SEPARATED FROM THE HOUSE AND STILL ATTACHED TO THE DRIVE WIRE, IN A TROMBONING FASHION. OUR ATTEMPTS TO COLLECT ADDITIONAL INFORMATION REGARDING PATIENT OUTCOME WERE UNSUCCESSFUL. WHILE THE COMPLAINANT DID NOT SPECIFY IF THE PATIENT EXPERIENCED ANY ADVERSE EFFECTS OR REQUIRED ADDITIONAL MEDICAL PROCEDURES DUE TO THIS EVENT, THE INFORMATION ABLE TO BE COLLECTED DOES NOT REASONABLY SUGGEST THE PATIENT WAS ADVERSELY IMPACTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1378541 | INSTINCT PLUS ENDOSCOPIC CLIPPING DEVICE | PKL, LIGATOR, HEMORRHOIDAL | PKL | COOK ENDOSCOPY | G58010 | W4453419 | 00827002580107 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |