FDA Adverse Event Malfunction Summary report: N

PLATE TCBS//TCBS 100 EA

MDR report key: 12485094 · Received September 16, 2021

Report

Report Number
1119779-2021-01535
Event Type
Malfunction
Date Received
September 16, 2021
Date of Event
August 20, 2021
Report Date
October 25, 2021
Manufacturer
BECTON, DICKINSON & CO. (SPARKS)
Product Code
JSJ
PMA / PMN Number
EXEMPT
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
JA
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

H.6. INVESTIGATION: THE DEFECT WAS NOT CONFIRMED BECAUSE THERE IS NO PICTURE AND RETURNED SAMPLE. PRODUCT EXPIRED ON JUN 29. BD WILL CONTINUE TO MONITOR FOR TREND. PLEASE EXPLAIN WHY "FAILURE TO INHIBIT" IS SELECTED: GENERALLY SPEAKING, THIS PRODUCT DOESN'T ALLOW TO GROWTH MANY KINDS OF BACTERIA DUE TO INGREDIENTS AND PH VALUE. SINCE IT'S ALREADY EXPIRED, THOSE FEATURES BECAME WEAKER OR WENT DOWN TO INSUFFICIENT LEVEL TO PREVENT GROWING BACTERIA. H3 OTHER TEXT : SEE H.10.

Description of Event or Problem · 0

IT WAS REPORTED THAT 20 PLATE TCBS//TCBS 100 EA EXPERIENCED BIOLOGICAL CONTAMINATION, AND PHYSICAL DEFORMITIES. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER: ACCORDING TO THE CUSTOMER¿S REPORT, AIR BUBBLES AND BACTERIAL GROWTH WERE FOUND IN THE MEDIA.

Additional Manufacturer Narrative · 1

THERE IS NO 510(K) FOR THIS DEVICE AS IT IS MANUFACTURED OUTSIDE THE US AND NOT SOLD IN THE US BUT IS CONSIDERED TO BE SUBSTANTIALLY SIMILAR TO THE LEGALLY U.S. MARKETED DEVICE BD BBL¿ TCBS AGAR, CATALOG NUMBER 221872 WITH 510K NUMBER EXEMPT. DATE OF EVENT: UNKNOWN. THE DATE RECEIVED BY MANUFACTURER HAS BEEN USED. A DEVICE EVALUATION IS ANTICIPATED, BUT HAS NOT YET BEGUN. UPON COMPLETION OF THE INVESTIGATION, A SUPPLEMENTAL REPORT WILL BE FILED.

Description of Event or Problem · 1

IT WAS REPORTED THAT 20 PLATE TCBS//TCBS 100 EA EXPERIENCED BIOLOGICAL CONTAMINATION, AND PHYSICAL DEFORMITIES. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER: ACCORDING TO THE CUSTOMER¿S REPORT, AIR BUBBLES AND BACTERIAL GROWTH WERE FOUND IN THE MEDIA.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1376638 PLATE TCBS//TCBS 100 EA NA JSJ BECTON, DICKINSON & CO. (SPARKS) 1062888

Patients

Seq Age Sex Outcome Treatment
1 Unknown