FDA Adverse Event Injury Summary report: N

GRAFTMASTER RX CORONARY STENT GRAFT SYSTEM

MDR report key: 12481718 · Received September 16, 2021

Report

Report Number
2024168-2021-08215
Event Type
Injury
Date Received
September 16, 2021
Date of Event
August 23, 2021
Report Date
September 16, 2021
Manufacturer
ABBOTT VASCULAR
Product Code
MAF
UDI-DI
08717648176340
PMA / PMN Number
H000001
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
VA, US
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

THE DEVICE WAS NOT RETURNED FOR ANALYSIS. A REVIEW OF THE LOT HISTORY RECORD IDENTIFIED NO MANUFACTURING NONCONFORMITIES ISSUED TO THE REPORTED LOT THAT WOULD HAVE CONTRIBUTED TO THIS EVENT. ADDITIONALLY, A REVIEW OF THE COMPLAINT HISTORY IDENTIFIED NO SIMILAR INCIDENTS FROM THIS LOT. THE INVESTIGATION WAS UNABLE TO DETERMINE A CONCLUSIVE CAUSE FOR THE REPORTED STENT DISLODGEMENT; HOWEVER, THE REPORTED TREATMENTS APPEAR TO BE RELATED TO THE OPERATIONAL CONTEXT OF THE PROCEDURE. THERE IS NO INDICATION OF A PRODUCT QUALITY ISSUE WITH RESPECT TO MANUFACTURE, DESIGN OR LABELING.

Description of Event or Problem · 1

IT WAS REPORTED THAT ON (B)(6) 2021, THE PATIENT PRESENTED WITH A SAPHENOUS VEIN GRAFT (SVG) TO THE LEFT ANTERIOR DESCENDING (LAD) PERFORATION. A 3.5X26MM AND 3.5X16MM GRAFTMASTER STENT WERE SUCCESSFULLY IMPLANTED, SEALING THE PERFORATION AT THAT SITE. THERE WAS NO DEVICE MALFUNCTION. A 2.80X16MM GRAFTMASTER (1012580-16, 0112541) STENT DELIVERY SYSTEM ALSO ADVANCED. DURING POSITIONING, NO UNUSUAL RESISTANCE WAS NOTED, HOWEVER, THE STENT DISLODGED AT THE LAD OSTIUM. ADDITIONAL DILATATION WAS PERFORMED, DEPLOYING THE 2.80X16MM GRAFTMASTER STENT AT THE OSTIAL LAD. THE PERFORATION HAD SEALED AND THERE WAS NO ADVERSE PATIENT SEQUELA. PER PHYSICIAN, THE GRAFTMASTER DEVICES DID NOT CAUSE OR CONTRIBUTE TO COMPLICATIONS OR ADVERSE EVENTS. NO ADDITIONAL INFORMATION WAS PROVIDED REGARDING THIS ISSUE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1381993 GRAFTMASTER RX CORONARY STENT GRAFT SYSTEM CORONARY STENT DELIVERY SYSTEM MAF ABBOTT VASCULAR 1012580-16 0112541 08717648176340

Patients

Seq Age Sex Outcome Treatment
1 77 YR Required Intervention 2 GRAFTMASTER STENTS