FDA Adverse Event Malfunction Summary report: N

SYRINGE 50ML LL CLEAR 14GA 1-1/4IN

MDR report key: 12481398 · Received September 16, 2021

Report

Report Number
3003152976-2021-00601
Event Type
Malfunction
Date Received
September 16, 2021
Date of Event
August 19, 2021
Report Date
October 20, 2021
Manufacturer
BECTON DICKINSON, S.A.
Product Code
FMF
PMA / PMN Number
NA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
GM
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

H6: INVESTIGATION SUMMARY PHOTOS RECEIVED FOR INVESTIGATION, UPON VISUAL INSPECTION IT CAN BE SEEN THE PICTURES WERE TAKEN WITH THE BISTER CLOSED. DAMAGE IN THE BARREL OF THE SYRINGE CAN BE NOTICED, ADDITIONALLY LOT NUMBER AND TOP WEB OF THE UNITARY PACKAGING ARE ALSO SHOWN IN THE PICTURES. POSSIBLE ROOT CAUSE IS ASSOCIATED WITH DAMAGE PRODUCED IN THE ASSEMBLY PROCESS. A DEVICE HISTORY REVIEW WAS CONDUCTED FOR LOT NUMBER 1912211. FINAL PRODUCTS IN THIS MANUFACTURING LINE, FOR THIS REFERENCE ARE SAMPLED AND THEY ARE SUBJECTED TO VISUAL AND FUNCTIONAL INSPECTIONS DURING THE DIFFERENT MANUFACTURING SUB-PROCESSES ACCORDING TO PROCEDURES.

Description of Event or Problem · 0

IT WAS REPORTED SYRINGE 50ML LL CLEAR 14GA 1-1/4IN WAS UNABLE TO CONTAIN BLOOD/MEDICATION DUE TO A CRACK OR HOLE IN THE SYRINGE. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER, TRANSLATED FROM GERMAN: "DEFECTIVE PRODUCTS DETECTED IN JULY 2021".

Additional Manufacturer Narrative · 1

DATE OF EVENT: UNKNOWN. THE DATE RECEIVED BY MANUFACTURER HAS BEEN USED FOR THIS FIELD. A DEVICE EVALUATION IS ANTICIPATED BUT HAS NOT YET BEGUN. UPON COMPLETION OF THE INVESTIGATION, A SUPPLEMENTAL REPORT WILL BE FILED.

Description of Event or Problem · 1

IT WAS REPORTED SYRINGE 50ML LL CLEAR 14GA 1-1/4IN WAS UNABLE TO CONTAIN BLOOD/MEDICATION DUE TO A CRACK OR HOLE IN THE SYRINGE. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER, TRANSLATED FROM (B)(6): "DEFECTIVE PRODUCTS DETECTED IN (B)(6) 2021"

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1377933 SYRINGE 50ML LL CLEAR 14GA 1-1/4IN PISTON SYRINGE FMF BECTON DICKINSON, S.A. 1912211

Patients

Seq Age Sex Outcome Treatment
1 Unknown