ARCHITECT TSH REAGENT KIT
Report
- Report Number
- 3005094123-2021-00176
- Event Type
- Malfunction
- Date Received
- September 16, 2021
- Date of Event
- September 9, 2021
- Report Date
- November 11, 2021
- Manufacturer
- A.I.D.D LONGFORD
- Product Code
- JLW
- UDI-DI
- 00380740014247
- PMA / PMN Number
- K983442
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- EI
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
- Health Professional
- Yes
Narratives
THE EVALUATION OF COMPLAINT DATA FOR THE PRODUCT AND LIKELY CAUSE ARCHITECT TSH REAGENT LOT 28075UD00 IDENTIFIED NORMAL COMPLAINT ACTIVITY AND THERE ARE NO TRENDS FOR THE PRODUCT RELATED TO PATIENT RESULTS. NO CUSTOMER RETURNS WERE AVAILABLE FOR EVALUATION. HISTORICAL PERFORMANCE IN THE FIELD OF REAGENT LOT USING WORLDWIDE DATA THROUGH ABBOTTLINK WAS EVALUATED. THE MEDIAN POPULATION RESULT FOR THE LOT IS WITHIN ESTABLISHED BASELINES AND COMPARABLE WITH ALL OTHER LOTS IN THE FIELD AND CONFIRMS NO SYSTEMIC ISSUE FOR THIS LOT. MANUFACTURING DOCUMENTATION FOR THE LIKELY CAUSE LOT DID NOT IDENTIFY ANY ISSUES ASSOCIATED WITH THE COMPLAINT ISSUE. THE PERFORMANCE OF THE LIKELY CAUSE LOT WAS INVESTIGATED BY COMPLETING A REVIEW FOR NON-CONFORMANCES, POTENTIAL NON-CONFORMANCES AND DEVIATIONS RELATED TO THE LIKELY CAUSE LOT. THIS REVIEW DID NOT IDENTIFY ANY NON-CONFORMANCES, POTENTIAL NON-CONFORMANCES OR DEVIATIONS. A REVIEW OF LABELING CONCLUDED THAT THE ISSUE IS SUFFICIENTLY ADDRESSED. BASED ON OUR INVESTIGATION NO SYSTEMIC ISSUE OR PRODUCT DEFICIENCY OF THE ARCHITECT TSH REAGENT LOT 28075UD00 WAS IDENTIFIED.
AN EVALUATION IS IN PROCESS. A FOLLOW UP REPORT WILL BE SUBMITTED WHEN THE EVALUATION IS COMPLETE.
THE CUSTOMER REPORTED A FALSELY DECREASED ARCHITECT TSH RESULT ON A PATIENT. RESULTS PROVIDED: 09SEP2021 SID (B)(6)= 0.0055 MIU/L, REPEATED ON ANOTHER ARCHITECT = 0.9 MIU/ML. OTHER RESULTS ALSO PROVIDED: FT4 = 61.77 PMOL/L, ANOTHER ARCHITECT = 12.5 / 12.2 PMOL/L LH = 5.22 / 11.96 IU/L FSH = 6.26 / 4.37 IU/L NO IMPACT TO PATIENT MANAGEMENT WAS REPORTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1378965 | ARCHITECT TSH REAGENT KIT | RADIOIMMUNOASSAY, THYROID-STIMULATING HORMONE | JLW | A.I.D.D LONGFORD | 7K62-30 | 28075UD00 | 00380740014247 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Unknown | ARC I2000SR INST, 03M74-02, (B)(4).| ARC I2000SR INST, 03M74-02, (B)(4).| ARC I2000SR INST, 03M74-02, ISR08024 |