FDA Adverse Event Injury Summary report: N

ENFLOW IV FLUID/BLOOD WARMER

MDR report key: 12478071 · Received September 15, 2021

Report

Report Number
3008959444-2021-00010
Event Type
Injury
Date Received
September 15, 2021
Date of Event
July 22, 2021
Report Date
August 16, 2021
Manufacturer
VYAIRE MEDICAL
Product Code
LGZ
UDI-DI
10190752125780
PMA / PMN Number
K130867
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
FI
Reporter Occupation
NURSE

Narratives

Additional Manufacturer Narrative · 1

VYAIRE AND HTI, CONDUCTED AN INVESTIGATION AND PERFORMED A TEST WITH THE ENFLOW TEST SETUP WHERE THE ORIGINAL ISSUE HAS BEEN OBSERVED. CONTROLLER DISPLAYS "NOT HEATING", HOWEVER AS SEEN IN THE THERMOMETER DISPLAY, WARMER IS CAPABLE TO WARM THE CARTRIDGE UP TO 39.2 °C.IT ONLY CANNOT SEND THE TEMPERATURE MEASUREMENT TO THE CONTROLLER. THE WARMERS RECEIVED IN 2021 HAVE PASSED FUNCTIONALITY TEST WITH THE OLDER CONTROLLER. WHEN FUNCTIONAL TESTER HAS BEEN REPLACED WITH A NEWER ONE, POOR CONNECTION ISSUE WAS OBSERVED AT HTI. THERE IS A PROBABILITY OF POOR CONNECTION AMONG THE WARMERS IN VYAIRE INVENTORY. POOR DATA CONNECTION IS CAUSED BY THE HUBBELL PLUG INNER DIAMETER BEING OUT OF NEMA STANDARD SPECIFICATION (4.19-4.32 MM), WHEN HUBBELL PLUG IS INSERTED IN THE HUBBELL RECEPTACLE. THIS ONLY IMPACTS THE DATA, NOT POWER. VYAIRE MEDICAL WILL SUBMIT A SUPPLEMENTAL REPORT IN ACCORDANCE WITH 21 CFR SECTION 803.56 IF ADDITIONAL INFORMATION BECOMES AVAILABLE.

Description of Event or Problem · 1

THE CUSTOMER REPORTED THAT THE STUDY NURSE WAS PREPARING THE ENFLOW DEVICES (WARMER, CARTRIDGE, AND CONTROLLER) WITH THE TRAINED ANESTHETIST NURSE BEFORE THE SURGERY. WHEN THE ENFLOW WARMER WITH THE CARTRIDGE WAS CONNECTED TO THE CONTROLLER, THE TEXT "NOT HEATING" WAS SHOWING ON THE CONTROLLER'S SCREEN. DESPITE THE TEXT, IT WAS STATED THAT THE WARMER FELT WARM. THE STUDY NURSE REPLACED THE WARMER, AND THE TEXT DID NOT REAPPEAR AGAIN. IT HAPPENED DURING PATIENT USE, BUT NO HARM IS ASSOCIATED WITH THE INCIDENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1375378 ENFLOW IV FLUID/BLOOD WARMER WARMER, THERMAL, INFUSION FLUID LGZ VYAIRE MEDICAL ENFLOW IV FLUID/BLOOD WARMER NOT APPLICABLE 10190752125780

Patients

Seq Age Sex Outcome Treatment
1 Required Intervention