CUP: VERSAFITCUP CC TRIO 01.26.45.0048 ACETABULAR SHELL CC TRIO DIAM 48
Report
- Report Number
- 3005180920-2021-00731
- Event Type
- Injury
- Date Received
- September 15, 2021
- Date of Event
- August 16, 2021
- Report Date
- September 15, 2021
- Manufacturer
- MEDACTA INTERNATIONAL SA
- Product Code
- LZO
- UDI-DI
- 07630030807756
- PMA / PMN Number
- K103352
- Adverse Event
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- SZ
- Reporter Occupation
- 003
Narratives
BATCH REVIEW PERFORMED ON 23 AUGUST 2021: LOT 2007993: (B)(4) ITEMS MANUFACTURED AND RELEASED ON 19-NOV-2020. EXPIRATION DATE: 2025-NOV-08. NO ANOMALIES FOUND RELATED TO THE PROBLEM. TO DATE, (B)(4) ITEMS OF THE SAME LOT HAVE BEEN SOLD WITHOUT ANY SIMILAR REPORTED EVENT. CLINICAL EVALUATION PERFORMED BY MEDICAL AFFAIRS MANAGER: HIP DISLOCATION 4 MONTHS AFTER PRIMARY IMPLANTATION. A FEMORAL REVISION WAS PERFORMED FEW WEEKS AFTER PRIMARY SURGERY DUE TO PERIPROSTHETIC FRACTURE. HIP DISLOCATION MAY BE DUE TO COMPONENT POSITIONING OR INSUFFICIENT SOFT TISSUE TENSIONING; IT CAN HARDLY BE CAUSED BY STANDARD DEVICES. OTHER DEVICES INVOLVED: BATCH REVIEW PERFORMED ON 13 SEPTEMBER 2021: STEM: AMISTEM P 01.18.401 AMISTEM-P STD STEM #1 (K173794) LOT 2005833: (B)(4) ITEMS MANUFACTURED AND RELEASED ON 01-OCT-2020. EXPIRATION DATE: 2025-SEPT-21. NO ANOMALIES FOUND RELATED TO THE PROBLEM. TO DATE, (B)(4) ITEMS OF THE SAME LOT HAVE BEEN SOLD WITHOUT ANY SIMILAR REPORTED EVENT.
THE PATIENT HAD PRIMARY SURGERY ON (B)(6) 2021. THE PATIENT HAD FIRST REVISION SURGERY DUE TO BONE FRACTURE ON (B)(6) 2021. THE FRACTURE WAS CAUSED BY PATIENT'S FALL. ONLY AMISTEM-P STEM AND COCR HEAD WERE REVISED. M-VIZION STEM AND NEW COCR HEAD IMPLANTED. ON (B)(6) 2021 LUXATION OCCURRED DUE TO TOO MUCH ANTEVERSION OF THE CUP. CUP, HEAD, AND LINER WERE REVISED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1372057 | CUP: VERSAFITCUP CC TRIO 01.26.45.0048 ACETABULAR SHELL CC TRIO DIAM 48 | HIP ACETABULAR SHELL | LZO | MEDACTA INTERNATIONAL SA | 01.26.45.0048 | 2007993 | 07630030807756 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Required Intervention |