FDA Adverse Event Malfunction Summary report: N

BD SYRINGE 0.3ML 30GA 8MM

MDR report key: 12468431 · Received September 14, 2021

Report

Report Number
1920898-2021-00985
Event Type
Malfunction
Date Received
September 14, 2021
Date of Event
August 17, 2021
Report Date
September 17, 2021
Manufacturer
BD MEDICAL - DIABETES CARE
Product Code
FMF
PMA / PMN Number
K024112
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
JA
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 0

H.6. INVESTIGATION: CUSTOMER RETURNED SEVERAL IMAGES OF TWO 0.3ML, 30 GAUGE, 8MM SYRINGES FROM LOT 0286309. THE SYRINGES HAD THEIR NEEDLE SHIELDS AND HUBS SEPARATE FROM THE BARRELS. THE HUBS HAVE BECOME LODGED INSIDE THE SHIELDS. THERE IS NO DAMAGE TO EITHER THE CONNECTORS AT THE DISTAL TIPS OF THE BARRELS OR THEIR RESPECTIVE NEEDLE HUBS. NO SIGNS OF USE AND NO OTHER DEFECTS FOUND. A REVIEW OF THE DEVICE HISTORY RECORD WAS COMPLETED FOR BATCH # 0286309 ALL INSPECTIONS WERE PERFORMED PER THE APPLICABLE OPERATIONS QC SPECIFICATIONS. THERE WERE ZERO (0) NOTIFICATIONS NOTED THAT PERTAINED TO THE COMPLAINT. BASED ON THE IMAGES RECEIVED, BD WAS ABLE TO CONFIRM THE CUSTOMER¿S INDICATED FAILURE OF NEEDLE HUB SEPARATION. CAPA#1630423 WAS INITIATED. H3 OTHER TEXT : SEE H.10.

Description of Event or Problem · 0

IT WAS REPORTED THAT 2 BD SYRINGE 0.3ML 30GA 8MM HUBS SEPARATED FROM THE DEVICE. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER : THE CUSTOMER REPORTED THAT THE NEEDLE WITH THE SHIED WAS SEPARATED FROM THE BARREL.

Additional Manufacturer Narrative · 1

A DEVICE EVALUATION IS ANTICIPATED, BUT HAS NOT YET BEGUN. UPON COMPLETION OF THE INVESTIGATION, A SUPPLEMENTAL REPORT WILL BE FILED.

Description of Event or Problem · 1

IT WAS REPORTED THAT 2 BD SYRINGE 0.3ML 30GA 8MM HUBS SEPARATED FROM THE DEVICE. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER : THE CUSTOMER REPORTED THAT THE NEEDLE WITH THE SHIED WAS SEPARATED FROM THE BARREL.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1367131 BD SYRINGE 0.3ML 30GA 8MM PISTON SYRINGE FMF BD MEDICAL - DIABETES CARE 0286309

Patients

Seq Age Sex Outcome Treatment
1