FDA Adverse Event Malfunction Summary report: N

I-STAT EC8+ CARTRIDGE

MDR report key: 12467358 · Received September 14, 2021

Report

Report Number
2245578-2021-00073
Event Type
Malfunction
Date Received
September 14, 2021
Date of Event
August 25, 2021
Report Date
October 18, 2021
Manufacturer
ABBOTT POINT OF CARE
Product Code
CHL
UDI-DI
10054749000071
PMA / PMN Number
K940918
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CA, US
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 0

APOC INCIDENT: # (B)(4). THE INVESTIGATION WAS COMPLETED ON 14-OCT-2021. A REVIEW OF THE DEVICE HISTORY RECORD (DHR) CONFIRMED THE CARTRIDGE LOT MET FINISHED GOODS (FG) RELEASE CRITERIA. RETAINED TESTING MET THE ACCEPTANCE CRITERIA FOUND IN Q04.01.003 REV. AG, APPENDIX 1 - PRODUCT COMPLAINT LEVEL 2 AND LEVEL 3 INVESTIGATION PROCEDURE. NO DEFICIENCY HAS BEEN DETERMINED FOR EC8+ CARTRIDGE LOT K21082.

Description of Event or Problem · 0

NA.

Additional Manufacturer Narrative · 1

APOC INCIDENT # (B)(4). APOC LABELING WILL BE EVALUATED DURING THE INVESTIGATION AS PERTAINING TO THE EVENT.

Description of Event or Problem · 1

ON (B)(6) 2021, ABBOTT POINT OF CARE WAS CONTACTED BY A CUSTOMER REGARDING I-STAT EC8+ CARTRIDGES THAT YIELDED A SUSPECTED DISCREPANT SODIUM RESULT OF 139 MMOL/L ON DOG. THERE WAS NO PATIENT INFORMATION AVAILABLE AT THE TIME OF THIS REPORT. METHOD, DATE, TESTED, RESULT (UNITS) SAMPLE: (B)(6). (B)(6): I-STAT REFERENCE RANGE FOR SODIUM (138-146 MMOL/L). AT THIS TIME THERE IS NO REASON TO SUSPECT A MALFUNCTION EXISTS. THE REPORTING DECISION WAS BASED ON THE LIMITED INFORMATION AVAILABLE THAT SUGGESTS THE PRODUCT WAS NOT PERFORMING WITHIN THE VARIABILITY OF THE ASSAY. CUSTOMER STATES THE PATIENT DIED ON (B)(6) 2021 AT 6:30PM DUE TO CARDIAC ARREST.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1367445 I-STAT EC8+ CARTRIDGE EC8+ CARTRIDGE CHL ABBOTT POINT OF CARE NA K21082 10054749000071

Patients

Seq Age Sex Outcome Treatment
1