INFANT CONTINUOUS FLOW BREATHING CIRCUIT
Report
- Report Number
- 9611451-2008-00683
- Event Type
- Malfunction
- Date Received
- November 21, 2008
- Date of Event
- October 22, 2008
- Report Date
- October 23, 2008
- Manufacturer
- FISHER & PAYKEL HEALTHCARE, LTD.
- Product Code
- BTT
- PMA / PMN Number
- K020332
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- SP
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
Narratives
THE MEDICAL DEVICES AND ACCESSORIES ASSOCIATED WITH THIS INCIDENT HAVE ONLY RECENTLY BEEN RETURNED TO FISHER & PAYKEL HEALTHCARE FOR INVESTIGATION. A THOROUGH INVESTIGATION IS CURRENTLY UNDERWAY. IN ORDER TO FURTHER ASSIST IN OUR ANALYSIS OF THE ROOT CAUSE OF THIS INCIDENT, FISHER & PAYKEL HEALTHCARE HAS MADE FURTHER INQUIRY IN RESPECT OF THE SURROUNDING THE EVENT, EQUIPMENT SET-UP AND USAGE. A TRANSLATION INTO THE ENGLISH LANGUAGE OF THE RESPONSES TO OUR QUESTIONS IS CURRENTLY BEING PREPARED AND THEREFORE, NOT YET AVAILABLE. WE WILL PROVIDE A FOLLOW-UP REPORT OUTLINING OUR ANALYSIS OF THE INCIDENT WHEN INVESTIGATION RESULTS BECOME AVAILABLE.
A HOSPITAL IN A FOREIGN COUNTRY REPORTED VIA A FISHER & PAYKEL HEALTHCARE REPRESENTATIVE AS FOLLOWS: A FIRE OCCURRED IN A SET-UP INVOLVING THE RT329 INFANT CONTINUOUS FLOW BREATHING CIRCUIT ('THE BREATHING CIRCUIT') AND MR290 AUTOFEED HUMIDIFICATION CHAMBER ('THE MR290 CHAMBER'). THE FIRE OCCURRED IN THE VICINITY OF THE MR290 CHAMBER PORT AND BREATHING CIRCUIT CONNECTOR. OTHER COMPONENTS OF THE BREATHING CIRCUIT SET-UP INCLUDED THE MR700 RESPIRATORY HUMIDIFIER, 900MR900 HEATER WIRE ADAPTOR (LOT NUMBER 061019 AND 900MR569 TEMPERATURE PROBE (LOT NUMBER 060831) AND HUDSON NASAL CANULA (REF 41828). THE SET-UP HAD BEEN IN USE ON A PATIENT UNDERGOING LOW FLOW OXYGEN THERAPY AT A FLOW RATE OF 0.2 TO 0.3 L/MIN FOR A PERIOD OF 30-45 MINUTES. NEITHER THE PATIENT NOR MEMBERS OF THE MEDICAL TEAM SUSTAINED ANY INJURIES AS A RESULT OF THIS INCIDENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | INFANT CONTINUOUS FLOW BREATHING CIRCUIT | BTT | FISHER & PAYKEL HEALTHCARE, LTD. | RT329 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | AUTOFEED HUMIDIFICATION CHAMBER| FISHER & PAYKEL HEALTHCARE'S MR290| 900MR900 HEATER WIRE ADAPTOR| 900MR569 TEMPERATURE PROBE| MR700 RESPIRATORY HUMIDIFIER| HUDSON NASAL CANULA |