FDA Adverse Event Malfunction Summary report: N

YC-200

MDR report key: 12465408 · Received September 13, 2021

Report

Report Number
0002936921-2021-00004
Event Type
Malfunction
Date Received
September 13, 2021
Date of Event
August 27, 2021
Report Date
September 13, 2021
Manufacturer
NIDEK CO. LTD.
Product Code
HQF
UDI-DI
04987669101452
PMA / PMN Number
K203130
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
MS, US
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 0

ON 09/02/2021, NIDEK INC FIELD SERVICE ENGINEER (FSE) WENT TO THE CUSTOMER SITE. THE REPORTED ISSUE WAS NOT DUPLICATED. FSE TESTED LASER FOCUS AND FOCUS SHIFT, AND FOUND BOTH WITHIN SPECIFICATION. TESTED LASER ENERGY AND FOUND ENERGY WITHIN SPECIFICATION. 10MJ=10MJ. TESTED LASER SLIT ILLUMINATION AND FOUND IT WITHIN SPECIFICATION. THE SYSTEM IS OPERATIONAL. THE ROOT CAUSE OF THE ISSUE IS INCONCLUSIVE. THE REPORTED ISSUE WAS NOT DUPLICATED.

Description of Event or Problem · 0

ON (B)(6) 2021, NIDEK INC CUSTOMER REPRESENTATIVE RECEIVED AN EMAIL FROM THE CUSTOMER TO REPORT THAT THE DOCTOR SAW PITTING ON IOL BUT PUSHED THE LASER FORWARD TO STOP IT FROM HAPPENING. THE DOCTOR REPORTS THAT IF FOCUSED BEHIND THE CAPSULE WITH THE AIMING BEAMS SPLIT, IT DOES FINE. THE DOCTOR INFORMED NIDEK INC THAT THE PATIENT IS FINE, PITTING WAS INFERIOR TO UNDILATED PUPIL AND WILL NOT INTERFERE WITH VISION. PATIENT WILL NOT HAVE ANY ADVERSE EFFECTS. NIDEK INC CONSIDERS THIS ISSUE ON YC-200 AS A REPORTABLE EVENT AS AN UNDESIRABLE CONDITION OCCURRED WHILE USING THE DEVICE. TECHNICAL BACKGROUND: THE OPHTHALMIC YAG LASER SYSTEM YC-200 CONSISTS OF A SLIT LAMP AND THE YAG LASER AND IS INDICATED FOR THE PERFORMANCE OF POSTERIOR CAPSULOTOMY, PUPILLARY MEMBRANECTOMY, IRIDOTOMY (HOLE IN THE IRIS) AND SELECTIVE LASER TRABECULOPLASTY. THIS DEVICE USES PASSIVE Q-SWITCHING FOR LASER OSCILLATION. WITH PASSIVE Q-SWITCHING, THE DEVICE FIRST ACCUMULATES SUFFICIENT ENERGY WITHIN THE LASER CAVITY THEN EMITS LASER PULSES OF HIGH PEAK POWER FOR EXTREMELY SHORT DURATIONS BY USE OF THE Q-SWITCHED OPTICAL ELEMENT WHICH CHANGES BETWEEN TRANSPARENT AND OPAQUE ACCORDING TO THE ACCUMULATION OF ENERGY.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1363197 YC-200 YAG LASER HQF NIDEK CO. LTD. 04987669101452

Patients

Seq Age Sex Outcome Treatment
1 Other