FDA Adverse Event Other Summary report: N

PRA INTERNAL ABUTMENT - LEVEL DRIVER, RATCHET

MDR report key: 1246352 · Received November 19, 2008

Report

Report Number
1060818-2008-00001
Event Type
Other
Date Received
November 19, 2008
Date of Event
October 28, 2008
Report Date
November 19, 2008
Manufacturer
BIOHORIZONS IMPLANT SYSTEMS, INC.
Product Code
DZJ
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
HI, US
Reporter Occupation
DENTIST

Narratives

Additional Manufacturer Narrative · 1

PRA WAS OVERLOADED CAUSING YIELDING AND FRACTURE OF THE TIP.

Description of Event or Problem · 1

PRA BROKE WHILE DRIVING THE IMPLANT INTO THE OSTEOTOMY. IMPLANT WAS REMOVED BECAUSE A FRAGMENT OF THE DRIVER COULD NOT BE REMOVED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 PRA INTERNAL ABUTMENT - LEVEL DRIVER, RATCHET PRA INTER. ABUTMENT - LEVEL DRIVER, RATCHET DZJ BIOHORIZONS IMPLANT SYSTEMS, INC.

Patients

Seq Age Sex Outcome Treatment
1 Required Intervention