FDA Adverse Event Malfunction Summary report: N

SYRINGE 30ML LL S/C 56

MDR report key: 12461072 · Received September 13, 2021

Report

Report Number
1911916-2021-00978
Event Type
Malfunction
Date Received
September 13, 2021
Date of Event
August 2, 2021
Report Date
September 20, 2021
Manufacturer
BD MEDICAL (BD WEST) MEDICAL SURGICAL
Product Code
FMF
UDI-DI
30382903028321
PMA / PMN Number
K980987
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CA
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 0

THE FOLLOWING FIELDS WERE UPDATED DUE TO ADDITIONAL INFORMATION: D10: DEVICE AVAILABLE FOR EVAL YES. D10: RETURNED TO MANUFACTURER ON: 2021-09-17. INVESTIGATION SUMMARY: IT WAS REPORTED THE DRUG LEAKED THROUGH THE PLUNGER. TO AID IN THE INVESTIGATION, ONE SAMPLE IN A SEALED PLASTIC BAG WAS RECEIVED FOR EVALUATION BY OUR QUALITY TEAM. A VISUAL INSPECTION WAS PERFORMED. THE SYRINGE HAS 25ML OF A CHEMO DRUG AND NO SOLUTION IS OBSERVED PAST THE STOPPER. NO OTHER DEFECTS OR IMPERFECTIONS WERE OBSERVED. A DEVICE HISTORY RECORD REVIEW WAS COMPLETED FOR PROVIDED MATERIAL NUMBER 302832, LOT NUMBER 1091323. THE REVIEW DID NOT REVEAL ANY DETECTED QUALITY ISSUES DURING THE PRODUCTION OF THIS LOT THAT COULD HAVE CONTRIBUTED TO THE REPORTED DEFECT. BASED ON THE INVESTIGATION AND WITH THE RETURNED SAMPLE ANALYSIS THE SYMPTOM REPORTED BY THE CUSTOMER COULD NOT BE CONFIRMED. H3 OTHER TEXT : SEE H10.

Description of Event or Problem · 0

IT WAS REPORTED WHEN USING THE SYRINGE 30ML LL S/C 56, THE DEVICE EXPERIENCED LEAKAGE PAST STOPPER/ PLUNGER. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER. THE CUSTOMER STATED: IT WAS REPORTED THAT DRUG LEAKED THROUGH THE PLUNGER.

Additional Manufacturer Narrative · 1

INVESTIGATION SUMMARY: A DEVICE HISTORY RECORD REVIEW WAS COMPLETED BY OUR QUALITY ENGINEER TEAM FOR PROVIDED MATERIAL NUMBER 302832 AND LOT NUMBER 1091323. THE REVIEW DID NOT REVEAL ANY DETECTED ABNORMALITIES DURING THE PRODUCTION PROCESS THAT COULD HAVE CONTRIBUTED TO THIS DEFECT AND ALL QUALITY TESTS WERE FOUND TO BE WITHIN SPECIFICATION. AS A SAMPLE WAS UNAVAILABLE FOR RETURN, A THOROUGH SAMPLE INVESTIGATION COULD NOT BE COMPLETED. BASED ON THE INVESTIGATION RESULTS, AN EXACT CAUSE FOR THIS INCIDENT COULD NOT BE IDENTIFIED. OUR QUALITY TEAM REGULARLY REVIEWS THE COLLECTED DATA FOR IDENTIFICATION OF EMERGING TRENDS.

Description of Event or Problem · 1

IT WAS REPORTED WHEN USING THE SYRINGE 30ML LL S/C 56, THE DEVICE EXPERIENCED LEAKAGE PAST STOPPER/ PLUNGER. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER. THE CUSTOMER STATED: IT WAS REPORTED THAT DRUG LEAKED THROUGH THE PLUNGER.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1359970 SYRINGE 30ML LL S/C 56 PISTON SYRINGE FMF BD MEDICAL (BD WEST) MEDICAL SURGICAL 302832 1091323 30382903028321

Patients

Seq Age Sex Outcome Treatment
1