COBAS EGFR MUTATION TEST V2
Report
- Report Number
- 2243471-2021-03272
- Event Type
- Malfunction
- Date Received
- September 13, 2021
- Date of Event
- April 16, 2021
- Report Date
- September 13, 2021
- Manufacturer
- ROCHE MOLECULAR SYSTEMS, INC. - BRANCHBURG
- Product Code
- OWD
- UDI-DI
- 00875197005448
- PMA / PMN Number
- P150047
- Removal / Correction Number
- 2243471-08-24-2021-003-C
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- SI
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
Narratives
ROCHE RECEIVED COMPLAINTS FROM CUSTOMERS REPORTING THE GENERATION OF FALSE MUTATION DETECTED RESULTS FOR THE EXON 20 INSERTION (EX20INS) MUTATION WHEN USING THE COBAS® EGFR MUTATION TEST V2. DURING IN-HOUSE TESTING USING CUSTOMER-PROVIDED FFPET SAMPLES, AN EX20INS FALSE MUTATION DETECTED RESULT WAS REPRODUCED FOR ONE OUT OF 8 FFPET SAMPLES, WHICH WAS PROCESSED FOLLOWING THE VALIDATED SAMPLE PREPARATION METHOD FROM THE INSTRUCTIONS FOR USE. ALTHOUGH THE MAJORITY OF CASES REPORTED WERE FROM USERS USING FFPET SAMPLES, THE GENERATION OF FALSE MUTATION DETECTED EX20INS RESULTS WITH PLASMA SPECIMENS WAS REPORTED IN ONE CASE. A FALSE MUTATION DETECTED EX20INS RESULT COULD LEAD TO HARM UNDER SPECIFIC SCENARIOS. CONSIGNEES WILL BE NOTIFIED OF THE ISSUE WITH INSTRUCTION TO FOLLOW THE INSTRUCTIONS FOR USE FOR SAMPLE INPUT REQUIREMENTS. ADDITIONALLY, IF AN EX20INS MUTATION DETECTED RESULT IS GENERATED WITH THE COBAS® EGFR MUTATION TEST V2, CUSTOMERS MUST CONFIRM THE RESULT WITH ANOTHER METHOD (E.G., SEQUENCING OR OTHER PCR-BASED TESTS). (B)(6). (B)(4).
A CUSTOMER IN SLOVENIA REPORTED THE GENERATION OF FALSE MUTATION DETECTED RESULTS FOR EXON20 INSERTION MUTATION FOR SEVERAL SAMPLES THAT WERE TESTED WITH THE COBAS EGFR MUTATION TEST V2 (LOTS G05117 AND G25391). SOME OF THE SAMPLE TYPES, SPECIFICALLY CYTOLOGICAL SAMPLES, WERE NOT VALIDATED FOR USE WITH THE COBAS EGFR MUTATION TEST, V2 AND FOR ON-LABEL SAMPLES (FFPET), THE SITE WAS USING MORE THAN ONE 5-MICRO FFPET SAMPLE DURING THEIR PREPARATION PROCESS. NO HARM OR INJURY WAS INDICATED. TWO MDRS WILL BE FILED, ONE FOR EACH ALLEGED KIT LOT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1359958 | COBAS EGFR MUTATION TEST V2 | SOMATIC GENE MUTATION DETECTION SYSTEM | OWD | ROCHE MOLECULAR SYSTEMS, INC. - BRANCHBURG | NA | G25391 | 00875197005448 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |