PLATE CDC ANAEROBE 5% SB 100 EA JP
Report
- Report Number
- 1119779-2021-01488
- Event Type
- Malfunction
- Date Received
- September 9, 2021
- Date of Event
- August 13, 2021
- Report Date
- August 30, 2021
- Manufacturer
- BECTON, DICKINSON & CO. (SPARKS)
- Product Code
- JSG
- PMA / PMN Number
- K803025
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- SN
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
Narratives
THERE IS NO 510(K) FOR THIS DEVICE AS IT IS MANUFACTURED OUTSIDE THE US AND NOT SOLD IN THE US BUT IS CONSIDERED TO BE SUBSTANTIALLY SIMILAR TO THE LEGALLY U.S. MARKETED DEVICE BD BBL¿ CDC ANAEROBE 5% SHEEP BLOOD AGAR CATALOG NUMBER 2221733 WITH 510K NUMBER K803025. INVESTIGATION SUMMARY: THE COMPLAINT WAS CONFIRMED AS THE REPORTED DEFECT ON A PICTURE. THE ISSUE WAS CONTAMINATION. WE OBSERVED SOME COLONIES ON THE AGAR BEDS. NO ISSUE IN BATCH HISTORY RECORD REVIEW AND QC ATTRIBUTE TEST THAT WE HELD. NO ISSUE IN RETENTION SAMPLE. WE COULD NOT DETERMINE ROOT CAUSE OF THIS ISSUE. NO TREND. WE WILL CONTINUE TO MONITOR THE TREND THIS ISSUE.
IT WAS REPORTED THAT PLATE CDC ANAEROBE 5% SB 100 EA (B)(4) WAS CONTAMINATED WITH BACTERIA COLONIES. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER: WOULD LIKE TO FEEDBACK THAT THERE ARE A COUPLE OF CDC PLATES ((B)(4) PLATES) WHICH ARE CONTAMINATED WITH BACTERIA COLONIES.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1344156 | PLATE CDC ANAEROBE 5% SB 100 EA JP | CULTURE MEDIA, NON-SELECTIVE AND NON-DIFFERENTIAL | JSG | BECTON, DICKINSON & CO. (SPARKS) | 1159074 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |