DREAMSTATION AUTO CPAP W/HUM/HT/CELL,DOM
Report
- Report Number
- 2518422-2021-03882
- Event Type
- Malfunction
- Date Received
- September 9, 2021
- Date of Event
- July 16, 2021
- Report Date
- July 6, 2023
- Manufacturer
- RESPIRONICS, INC.
- Product Code
- BZD
- PMA / PMN Number
- K131982
- Removal / Correction Number
- RES 88058
- Adverse Event
- Yes
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- US
- Reporter Occupation
- 003
Narratives
MDR 2518422-2021-03882 IS THE ORIGINAL REPORT ASSOCIATED WITH THIS DEVICE. A DUPLICATE REPORT HAS BEEN SUBMITTED FOR THIS DEVICE UNDER REPORT MDR 2518422-2021-04049. THIS CORRECTION REPORT IS BEING FILED FOR AWARENESS OF THIS DUPLICATE REPORT. A CORRECTION REPORT 2518422-2021-04049-1 HAS BEEN FILED FOR AWARENESS ABOUT THE DUPLICATE REPORT.
THE MANUFACTURER PREVIOUSLY REPORTED AN ALLEGATION OF AN ISSUE RELATED TO SOUND ABATEMENT FOAM. ADDITIONAL INFORMATION WAS RECEIVED AND SECTION B5 SHOULD BE REPORTED AS: THE MANUFACTURER WAS CONTACTED IN REFERENCE TO THE VOLUNTARY FIELD SAFETY NOTICE / RECALL NOTIFICATION RELATED TO THE SOUND ABATEMENT FOAM IN CERTAIN CPAP, BIPAP, AND MECHANICAL VENTILATOR DEVICES. THE MANUFACTURER PREVIOUSLY RECEIVED INFORMATION ALLEGING A CONTINUOUS POSITIVE AIRWAY PRESSURE (CPAP) DEVICE'S SOUND ABATEMENT FOAM BECAME DEGRADED AND CAUSED A PATIENT DEVELOP RESPIRATORY ISSUES AND MUCOUS. THE REPORTED EVENT OF RESPIRATORY ISSUES, MUCOUS AND ITS REPORTED SEVERITY WAS REVIEWED BY THE MANUFACTURE'S CLINICAL EXPERT. THIS EVENT IS ASSESSED AS NOT RELATED TO THE DEVICE IN THIS CASE. BASED ON THE AVAILABLE INFORMATION, THE MANUFACTURE CONCLUDES NO FURTHER ACTION IS NECESSARY. THE MEDICAL INTERVENTION RECEIVED BY THE PATIENT IN THE FORM OF PRESCRIBED MEDICATION. THE DEVICE HAS NOT YET RETURNED TO THE MANUFACTURER FOR EVALUATION. HENCE, NO FURTHER INVESTIGATION CAN BE PERFORMED. IF ANY ADDITIONAL INFORMATION IS RECEIVED, A FOLLOW UP REPORT WILL BE FILED. SECTION(S) B1, B2, HAS CHANGED RELATED TO THE COMPLAINT CHANGING FROM THE REPORTED ADVERSE EVENT TO A PRODUCT PROBLEM. SECTION H1 HAS CHANGED TO REFLECT A MALFUNCTION. SECTION H6 HEALTH EFFECT- IMPACT CODE, TYPE OF INVESTIGATION FINDINGS AND INVESTIGATION CONCLUSIONS HAS BEEN UPDATED.
THE MANUFACTURER RECEIVED INFORMATION ALLEGING A CONTINUOUS POSITIVE AIRWAY PRESSURE (CPAP) DEVICE'S SOUND ABATEMENT FOAM BECAME DEGRADED AND CAUSED THE PATIENT TO DEVELOP RESPIRATORY ISSUES. THE PATIENT WAS PRESCRIBED MEDICATION IN RESPONSE TO THE REPORTED EVENT. THIS ISSUE WAS REPORTED TO THE FDA PER 21 CFR 806. THE DEVICE WILL BE CORRECTED PER RES 88058.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1343554 | DREAMSTATION AUTO CPAP W/HUM/HT/CELL,DOM | VENTILATOR, NON-CONTINUOUS (RESPIRATOR) | BZD | RESPIRONICS, INC. | DSX500T11C |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Unknown | Required Intervention |