GORE® VIABAHN® ENDOPROSTHESIS WITH HEPARIN BIOACTIVE SURFACE
Report
- Report Number
- 2017233-2021-02345
- Event Type
- Injury
- Date Received
- September 9, 2021
- Date of Event
- October 22, 2020
- Report Date
- March 17, 2022
- Manufacturer
- W. L. GORE & ASSOCIATES, INC.
- Product Code
- NIP
- PMA / PMN Number
- P040037
- Adverse Event
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- JA
- Reporter Occupation
- PHYSICIAN
- Health Professional
- Yes
Narratives
UPDATED ATTACHMENTS. ATTACHED REVIEWED LITERATURE ARTICLE, "A CASE OF STENT-GRAFT IMPLANTATION FOR POSTPANCREATICODUODENECTOMY HEMORRHAGE IN A PATIENT WITH A RECONSTRUCTED GASTRIC TUBE" (B)(6), MD, ET AL. RADIOLOGY CASE REPORTS. 2020 DEC; 15(12): 2710¿2713. PUBLISHED ONLINE 2020 OCT 22.
AS THE (DEVICES) WHERE OVERLAPPED AND IT IS UNKNOWN WHICH DEVICE(S) MAY HAVE CONTRIBUTED TO THE THROMBOSIS/OCCLUSION/RESTENOSIS, AN ADDITIONAL ( DEVICE ) HAVE BEEN INCLUDED IN THIS REPORT: BOTH GORE® VIABAHN® ENDOPROSTHESIS DEVICES WERE PLACED IN SAME ANATOMICAL LOCATION AT THE SAME TIME. LOT NUMBER AND UDI NUMBER ARE UNKNOWN. LOT NUMBER REQUESTED BUT NOT MADE AVAILABLE. CBAS® HEPARIN SURFACE INCORPORATES CBAS-HEPARIN MANUFACTURED FROM HEPARIN SODIUM APL, WHICH IN COVALENTLY BOUND TO THE DEVICE SURFACE AND IS ESSENTIALLY/NON-ELUTING.
THE FOLLOWING LITERATURE WAS REVIEWED. "A CASE OF STENT-GRAFT IMPLANTATION FOR POSTPANCREATICODUODENECTOMY HEMORRHAGE IN A PATIENT WITH A RECONSTRUCTED GASTRIC TUBE" TOMOHIRO KOMADA, MD, ET AL. RADIOLOGY CASE REPORTS. 2020 DEC; 15(12): 2710¿2713. PUBLISHED ONLINE 2020 OCT 22. IN PATIENTS WITH A RECONSTRUCTED GASTRIC TUBE, THE RIGHT GASTROEPIPLOIC ARTERY IS A VERY IMPORTANT FEEDING ARTERY OF THE TUBE, WHICH MUST BE PRESERVED WHEN PERFORMING A PANCREATICODUODENECTOMY. A (B)(6) MAN WITH A RECONSTRUCTED GASTRIC TUBE UNDERWENT PANCREATICODUODENECTOMY FOR DISTAL BILE DUCT CARCINOMA. ON POSTOPERATIVE DAY 8, HE HAD AN ARTERIAL HEMORRHAGE FROM A DRAIN, APPARENTLY FROM A LIGATION OF THE ANTERIOR SUPERIOR DUODENAL ARTERY. HE, THEREFORE, UNDERWENT STENT-GRAFT (6 MM × 2.5 CM, VIABAHN; W. L. GORE, NEWARK, DE) PLACEMENT IN THE GASTRODUODENAL ARTERY ON POSTOPERATIVE DAY 16. ANGIOGRAPHY AFTER IMPLANTATION DID NOT DELINEATE THE PROXIMAL PORTION OF THE STENT-GRAFT, AND THE POSSIBILITY OF DISSECTION OF THE PROXIMAL VESSEL WAS CONSIDERED. A 6 MM × 5 CM STENT-GRAFT (VIABAHN) WAS THEN PLACED IN FRONT OF THE FIRST STENT-GRAFT IN THE STENT, AND FURTHER LIGHT CRIMPING WAS PERFORMED WITH A BALLOON CATHETER (SABER; CORDIS, CARDINAL HEALTH, DUBLIN, OH). TWO DAYS AFTER STENT-GRAFT IMPLANTATION, THE STENT-GRAFTS WERE TEMPORARILY OCCLUDED, AND THE GASTRIC TUBE WAS NECROTIZING. HOWEVER, THROMBOLYTIC THERAPY ALLOWED THE STENT-GRAFTS TO REOPEN AND PREVENTED GASTRIC TUBE NECROSIS. THE ISCHEMIC CHANGES IN THE GASTRIC TUBE IMPROVED ON ENDOSCOPY 6 DAYS AFTER THE THROMBECTOMY (FIG. 3D). HEPARIN WAS CHANGED TO ANTIPLATELET MEDICATIONS (ASPIRIN AND CLOPIDOGREL SULFATE), AND THE PATIENT WAS DISCHARGED 53 DAYS AFTER SURGERY. THE CT AT 3 MONTHS AFTER STENT-GRAFT IMPLANTATION SHOWED THAT THE STENT-GRAFTS WERE OCCLUDED, BUT THE CONTRAST EFFECT OF THE GASTRIC TUBE WAS GOOD. THE PATIENT DID NOT HAVE ANY SYMPTOMS. WE BELIEVE OUR CASE OF STENT-GRAFT IMPLANTATION IN THE GASTRODUODENAL ARTERY IS THE FIRST OF THIS KIND TO SUCCESSFULLY PREVENT LETHAL NECROSIS OF THE GASTRIC TUBE.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1343342 | GORE® VIABAHN® ENDOPROSTHESIS WITH HEPARIN BIOACTIVE SURFACE | STENT, SUPERFICIAL FEMORAL ARTERY | NIP | W. L. GORE & ASSOCIATES, INC. |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 76 YR | Male | Other |