THUNDERBEAT OPEN FINE JAW
Report
- Report Number
- 8010047-2021-11458
- Event Type
- Malfunction
- Date Received
- September 8, 2021
- Date of Event
- August 19, 2021
- Report Date
- September 9, 2021
- Manufacturer
- OLYMPUS MEDICAL SYSTEMS CORP.
- Product Code
- GEI
- PMA / PMN Number
- K211838
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- JA
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
Narratives
THE SUBJECT DEVICE WAS RETURNED TO OLYMPUS MEDICAL SYSTEMS CORP. (OMSC) FOR EVALUATION. THE PROBE WAS BROKEN OFF AT 17 MM FROM THE DISTAL END. THERE WAS SCRATCH AROUND THE BROKEN POINT. THE FRACTURE SURFACE OF THE PROBE SHOWED THAT CRACK DEVELOPED. THE MANUFACTURING RECORD WAS REVIEWED AND FOUND NO IRREGULARITIES. BASED ON THE PAST SIMILAR CASES, IT WAS KNOWN THAT THE PROBE WAS CRACKED WHEN THE USER ACTIVATED OUTPUT CONTACTING WITH METAL OR SOMETHING HARD OBJECT SUCH AS METAL CLIPS, STAPLER, OR OTHER INSTRUMENTS, BESIDES THE PROBE WAS BROKEN OFF WHEN THE PROBE WAS SUBJECTED TO A LOAD. THE ABOVE DEVICE HANDLING HAS WARNED IN THE INSTRUCTION MANUAL.
OLYMPUS MEDICAL SYSTEMS CORP. (OMSC) WAS INFORMED BY THE HEALTH PROFESSIONAL THAT DURING THE PROCEDURE FOR HEAD AND NECK CANCER USING THE SUBJECT DEVICE, THE FOLLOWING EVENT OCCURRED. THE PROBE WAS BROKEN OFF. THE INTENDED PROCEDURE WAS COMPLETED WITH ANOTHER DEVICE. THERE WAS NO PATIENT INJURY REPORTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1339629 | THUNDERBEAT OPEN FINE JAW | ULTRASONIC SURGICAL DEVICE | GEI | OLYMPUS MEDICAL SYSTEMS CORP. | TB-0009OF | 16K |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |