TRUE METRIX AIR
Report
- Report Number
- 1000113657-2021-00543
- Event Type
- Malfunction
- Date Received
- September 8, 2021
- Date of Event
- August 13, 2021
- Report Date
- November 1, 2021
- Manufacturer
- TRIVIDIA HEALTH INC
- Product Code
- NBW
- UDI-DI
- 00021292009120
- PMA / PMN Number
- K150052
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- SC, US
- Reporter Occupation
- OTHER
Narratives
SECTIONS WITH ADDITIONAL INFORMATION AS OF 01-NOV-2021: H6: UPDATED FDA¿S TYPE, FINDINGS AND CONCLUSIONS CODES. H10: METER WAS NOT RETURNED FOR EVALUATION. TEST STRIPS WERE NOT RETURNED FOR EVALUATION. RETENTION TESTING WAS PERFORMED USING TEST STRIPS FROM THE SAME LOT. RETENTION STRIP LOT TESTED WITHIN SPECIFICATIONS. MOST LIKELY UNDERLYING ROOT CAUSE: MLC-058: USER HAD AN INACCURATE REFERENCE: SELF: THE PERSON IS USING THEMSELVES AS THE REFERENCE OR HOW THEY FEEL AT THE TIME THEY RUN THE BLOOD TEST.
INTERNAL REPORT REFERENCE NUMBER: (B)(4). METER AND TEST STRIPS WERE NOT RETURNED FOR EVALUATION. NOTE: MANUFACTURER CONTACTED CUSTOMER IN SEVERAL FOLLOW-UP CALLS TO ENSURE THE REPLACEMENT PRODUCTS RESOLVED THE INITIAL CONCERN - UNABLE TO ESTABLISH CONTACT WITH CUSTOMER AT THIS TIME.
CONSUMER REPORTED COMPLAINT FOR HIGH AND ERRATIC BLOOD GLUCOSE TEST RESULTS. THE CUSTOMER IS CONCERNED WITH TEST RESULTS FROM RESULTS OBTAINED OF 231, 233 AND 214 MG/DL. THE CUSTOMER¿S EXPECTED FASTING BLOOD GLUCOSE TEST RESULT RANGE IS 100-120 MG/DL. THE CUSTOMER FEELS WELL AND DID NOT REPORT ANY SYMPTOMS. MEDICAL ATTENTION IS NOT REPORTED AS A RESULT OF THE ACTUAL BLOOD GLUCOSE RESULTS. DURING THE CALL, A BACK TO BACK BLOOD TEST WAS NOT PERFORMED BY THE CUSTOMER. THE PRODUCT IS STORED ACCORDING TO SPECIFICATION IN THE BEDROOM. THE TEST STRIP LOT MANUFACTURER¿S EXPIRATION DATE IS 11/30/2022 AND TEST STRIPS WERE OPENED 2-3 DAYS PRIOR TO CALL. THE METER MEMORY WAS REVIEWED FOR PREVIOUS TEST RESULT HISTORY (ONLY 3 RESULTS PROVIDED): (B)(6).
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1341195 | TRUE METRIX AIR | SYSTEM, TEST BLOOD GLUCOSE, OVER THE COUNTER | NBW | TRIVIDIA HEALTH INC | STRIP, WGN TMX 30CT 24/CASE MG/DL | ZY4376S | 00021292009120 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |