IMPL TAPERED SCR-V SBM 3. 7MM 3.5MM 16MM
Report
- Report Number
- 0002023141-2021-02475
- Event Type
- Injury
- Date Received
- September 7, 2021
- Date of Event
- October 9, 2020
- Report Date
- January 4, 2022
- Manufacturer
- ZIMMER DENTAL
- Product Code
- DZE
- UDI-DI
- 00889024019539
- PMA / PMN Number
- K061410
- Adverse Event
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- CH
- Reporter Occupation
- 003
Narratives
THIS REPORT IS BEING SUBMITTED TO RELAY ADDITIONAL INFORMATION AND DEVICE EVALUATION. THE FOLLOWING SECTIONS ARE BEING REPORTED: H6: TYPE OF INVESTIGATION CODE WAS ADDED: 3331. DHR REVIEW WAS COMPLETED FOR THE SUBJECT LOT NUMBER (1231102). IT WAS CONFIRMED THAT ALL OPERATIONS AND INSPECTIONS WERE EXECUTED AS PER APPLICABLE PROCEDURE. NO DEVIATIONS OR NON-CONFORMANCES, WHICH COULD HAVE CAUSED OR CONTRIBUTED TO THE REPORTED EVENT WERE NOTED AS PART OF THE DHR. LOT WAS INSPECTED AND PASSED ALL ACCEPTANCE CRITERIA BY QA. STERILIZATION RECORD (OP150) WAS REVIEWED AND VERIFIED TO HAVE PASSED ALL STERILIZATION ACTIVITIES WITH NO ISSUES OR NONCONFORMITIES IDENTIFIED. COMPLAINT HISTORY REVIEW WAS PERFORMED FOR THE REPORTED LOT NUMBERS (1231102) FOR SIMILAR EVENTS (COMPLAINT CATEGORY KEYWORD: MEDICAL: INFECTION) AND NO OTHER COMPLAINT WAS IDENTIFIED. AS DOCUMENTED IN THE SUMMARY INVESTIGATION, CONTRIBUTING FACTORS FOR THE REPORTED EVENT LIKELY EXIST OUTSIDE OF ZIMMER BIOMET CONTROL, INCLUDING THOSE RELATED TO PATIENT BIOLOGICAL FACTORS/CONDITION AND SURGICAL TECHNIQUE. BASED ON THE SUMMARY INVESTIGATION, NO MALFUNCTION OCCURRED UPON INVESTIGATION. THE REPORTED EVENT REMAINS NON-VERIFIABLE AS IT IS A MEDICAL CONDITION.
NO ADDITIONAL EVENT INFORMATION AT THE TIME OF THIS REPORT.
ZIMMER BIOMET COMPLAINT NUMBER (B)(4). PATIENT IDENTIFIER- (B)(6). ADDITIONAL PMA/510(K) NUMBER ¿ K011028/K013227. A SUMMARY INVESTIGATION HAS BEEN COMPLETED FOR INFECTION EVENTS IDENTIFYING THAT A DEFINITIVE ROOT CAUSE CANNOT BE IDENTIFIED DUE TO THE WIDE RANGE OF EXTERNAL (NON-DESIGN/ NON-MANUFACTURING RELATED) FACTORS POTENTIALLY IMPACTING THE STERILITY OF THE IMPLANT AND ITS ENVIRONMENT. SHOULD ADDITIONAL INFORMATION BE RECEIVED WHICH INDICATES THAT THE DEVICE MAY HAVE CAUSED OR CONTRIBUTED TO THE INFECTION EVENT, AN ADDITIONAL REPORT WILL BE SUBMITTED.
THE DOCTOR INDICATED INFECTION. PAIN AT THE IMPLANT SITE WAS FOUND 3 MONTHS AFTER IMPLANTATION. THE PATIENT CAME TO THE CLINIC FOR EXAMINATION. THE IMPLANT WAS REMOVED FOR ANTI-INFLAMMATORY TREATMENT. TOOTH SITE #21.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1330461 | IMPL TAPERED SCR-V SBM 3. 7MM 3.5MM 16MM | DENTAL IMPLANT | DZE | ZIMMER DENTAL | TSVB16 | 1231102 | 00889024019539 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 59 YR | Male | Required Intervention |