FDA Adverse Event Injury Summary report: N

IMPL TAPERED SCR-V SBM 3. 7MM 3.5MM 16MM

MDR report key: 12435701 · Received September 7, 2021

Report

Report Number
0002023141-2021-02475
Event Type
Injury
Date Received
September 7, 2021
Date of Event
October 9, 2020
Report Date
January 4, 2022
Manufacturer
ZIMMER DENTAL
Product Code
DZE
UDI-DI
00889024019539
PMA / PMN Number
K061410
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
CH
Reporter Occupation
003

Narratives

Additional Manufacturer Narrative · 0

THIS REPORT IS BEING SUBMITTED TO RELAY ADDITIONAL INFORMATION AND DEVICE EVALUATION. THE FOLLOWING SECTIONS ARE BEING REPORTED: H6: TYPE OF INVESTIGATION CODE WAS ADDED: 3331. DHR REVIEW WAS COMPLETED FOR THE SUBJECT LOT NUMBER (1231102). IT WAS CONFIRMED THAT ALL OPERATIONS AND INSPECTIONS WERE EXECUTED AS PER APPLICABLE PROCEDURE. NO DEVIATIONS OR NON-CONFORMANCES, WHICH COULD HAVE CAUSED OR CONTRIBUTED TO THE REPORTED EVENT WERE NOTED AS PART OF THE DHR. LOT WAS INSPECTED AND PASSED ALL ACCEPTANCE CRITERIA BY QA. STERILIZATION RECORD (OP150) WAS REVIEWED AND VERIFIED TO HAVE PASSED ALL STERILIZATION ACTIVITIES WITH NO ISSUES OR NONCONFORMITIES IDENTIFIED. COMPLAINT HISTORY REVIEW WAS PERFORMED FOR THE REPORTED LOT NUMBERS (1231102) FOR SIMILAR EVENTS (COMPLAINT CATEGORY KEYWORD: MEDICAL: INFECTION) AND NO OTHER COMPLAINT WAS IDENTIFIED. AS DOCUMENTED IN THE SUMMARY INVESTIGATION, CONTRIBUTING FACTORS FOR THE REPORTED EVENT LIKELY EXIST OUTSIDE OF ZIMMER BIOMET CONTROL, INCLUDING THOSE RELATED TO PATIENT BIOLOGICAL FACTORS/CONDITION AND SURGICAL TECHNIQUE. BASED ON THE SUMMARY INVESTIGATION, NO MALFUNCTION OCCURRED UPON INVESTIGATION. THE REPORTED EVENT REMAINS NON-VERIFIABLE AS IT IS A MEDICAL CONDITION.

Description of Event or Problem · 0

NO ADDITIONAL EVENT INFORMATION AT THE TIME OF THIS REPORT.

Additional Manufacturer Narrative · 1

ZIMMER BIOMET COMPLAINT NUMBER (B)(4). PATIENT IDENTIFIER- (B)(6). ADDITIONAL PMA/510(K) NUMBER ¿ K011028/K013227. A SUMMARY INVESTIGATION HAS BEEN COMPLETED FOR INFECTION EVENTS IDENTIFYING THAT A DEFINITIVE ROOT CAUSE CANNOT BE IDENTIFIED DUE TO THE WIDE RANGE OF EXTERNAL (NON-DESIGN/ NON-MANUFACTURING RELATED) FACTORS POTENTIALLY IMPACTING THE STERILITY OF THE IMPLANT AND ITS ENVIRONMENT. SHOULD ADDITIONAL INFORMATION BE RECEIVED WHICH INDICATES THAT THE DEVICE MAY HAVE CAUSED OR CONTRIBUTED TO THE INFECTION EVENT, AN ADDITIONAL REPORT WILL BE SUBMITTED.

Description of Event or Problem · 1

THE DOCTOR INDICATED INFECTION. PAIN AT THE IMPLANT SITE WAS FOUND 3 MONTHS AFTER IMPLANTATION. THE PATIENT CAME TO THE CLINIC FOR EXAMINATION. THE IMPLANT WAS REMOVED FOR ANTI-INFLAMMATORY TREATMENT. TOOTH SITE #21.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1330461 IMPL TAPERED SCR-V SBM 3. 7MM 3.5MM 16MM DENTAL IMPLANT DZE ZIMMER DENTAL TSVB16 1231102 00889024019539

Patients

Seq Age Sex Outcome Treatment
1 59 YR Male Required Intervention