PUMP MMT-1780KPK 670G PATHWAY BLACK MG
Report
- Report Number
- 2032227-2021-189936
- Event Type
- Malfunction
- Date Received
- September 7, 2021
- Date of Event
- August 15, 2021
- Report Date
- February 9, 2022
- Manufacturer
- MEDTRONIC PUERTO RICO OPERATIONS CO.
- Product Code
- OZP
- UDI-DI
- 000000763000283520
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- IL, US
- Reporter Occupation
- OTHER
- Health Professional
- N
Narratives
RETAINER RING=BLACK. CUSTOMER COMPLAINED ON (B)(6) 2021 THE BUTTONS ARE NOT FUNCTIONING PROPERLY AND SCRATCHED LIQUID CRYSTAL DISPLAY WINDOW. THE INSULIN PUMP PASSED SELF TEST AND DISPLACEMENT TEST. ALL BUTTONS FUNCTION PROPERLY. DEVICE PASSED THE KEYPAD VOLTAGE TEST. NO DAMAGE NOTED TO KEYPAD ASSEMBLY AND CONNECTOR ON ELECTRICAL BOARD 1 WAS LOCKED PROPERLY DURING VISUAL INSPECTION. DEVICE SUCCESSFULLY DOWNLOADED TO THUS. THE ELECTRONIC ASSEMBLIES WERE INSPECTED AND NO ANOMALIES NOTED. THE FOLLOWING WERE NOTED DURING VISUAL INSPECTION: SCRATCHED CASE, PILLOWING KEYPAD OVERLAY AND SCRATCHED LIQUID CRYSTAL DISPLAY WINDOW. NO CRACKS ON CASE OR LIQUID CRYSTAL DISPLAY WINDOW NOTED. THE P-CAP OR RESERVOIR DOES LOCK PROPERLY. CONFIRMED ALL BUTTONS FUNCTIONED PROPERLY DURING TEST AND CONFIRMED SCRATCHED LIQUID CRYSTAL DISPLAY WINDOW. THIS MDR RELATED TO THE PUERTO RICO MANUFACTURING SITE HAS BEEN ASSIGNED A MEDWATCH NUMBER FROM THE MEDTRONIC MINIMED NORTHRIDGE SITE, PER VARIANCE 5. MEDTRONIC, INC. (MEDTRONIC) IS SUBMITTING THIS REPORT TO COMPLY WITH 21 C.F.R. PART 803, THE MEDICAL DEVICE REPORTING REGULATION. THIS REPORT IS BASED UPON INFORMATION OBTAINED BY MEDTRONIC, WHICH THE COMPANY MAY NOT HAVE BEEN ABLE TO FULLY INVESTIGATE OR VERIFY PRIOR TO THE DATE THE REPORT WAS REQUIRED BY THE FDA. MEDTRONIC HAS MADE REASONABLE EFFORTS TO OBTAIN MORE COMPLETE INFORMATION IN THE TIME ALLOTTED AND HAS PROVIDED AS MUCH INFORMATION AS IS AVAILABLE TO THE COMPANY AS OF THE SUBMISSION DATE THIS REPORT. THIS REPORT DOES NOT CONSTITUTE AN ADMISSION OR A CONCLUSION BY FDA, MEDTRONIC, OR ITS EMPLOYEES THAT THE DEVICE, MEDTRONIC, OR ITS EMPLOYEES CAUSED OR CONTRIBUTED TO THE EVENT DESCRIBED IN THE REPORT. IN PARTICULAR, THIS REPORT DOES NOT CONSTITUTE AN ADMISSION BY ANYONE THAT THE PRODUCT DESCRIBED IN THIS REPORT HAS ANY "DEFECTS" OR HAS "MALFUNCTIONED". THESE WORDS ARE INCLUDED IN THE FDA 3500A FORM AND ARE FIXED ITEMS FOR SELECTION CREATED BY THE FDA, TO CATEGORIZE THE TYPE OF EVENT SOLELY FOR THE PURPOSE OF REPORTING PURSUANT TO PART 803. MEDTRONIC OBJECTS TO THE USE OF THESE WORDS AND OTHERS LIKE IT BECAUSE OF THE LACK OF DEFINITION AND THE CONNOTATIONS IMPLIED BY THESE TERMS. THIS STATEMENT SHOULD BE INCLUDED WITH ANY INFORMATION OR REPORT DISCLOSED TO THE PUBLIC UNDER THE FREEDOM OF INFORMATION ACT.
(B)(4). CURRENTLY IT IS UNKNOWN WHETHER OR NOT THE DEVICE MAY HAVE CAUSED OR CONTRIBUTED TO THE EVENT AS NO PRODUCT HAS BEEN RETURNED. THE DEVICE WILL BE RETURNED FOR ANALYSIS AND FURTHER INFORMATION WILL FOLLOW ONCE THE ANALYSIS HAS BEEN COMPLETED. NO CONCLUSION CAN BE DRAWN AT THIS TIME.
INFORMATION RECEIVED BY MEDTRONIC INDICATED THAT THE INSULIN PUMP HAD KEYPAD ANOMALY. CUSTOMER STATED DOWN BUTTON SEEMED STUCK AND CUSTOMER WAS HAVING HARD TIME PRESSING DOWN BUTTON. CUSTOMER STATED THEY DID NOT RECEIVE STUCK BUTTON ALARM. CUSTOMER REPORTED CRACK AT FRONT OF PUMP ABOVE THE GRAPH SCREEN. CUSTOMER STATED THAT NUMBERS WERE RAMPING ON THEIR OWN. CUSTOMER STATED THE SCROLL BAR WAS MOVING WITH NO INPUT. TROUBLESHOOTING WAS PERFORMED. NO HARM REQUIRING MEDICAL INTERVENTION WAS REPORTED. THE INSULIN PUMP WILL BE RETURNED FOR ANALYSIS.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1334228 | PUMP MMT-1780KPK 670G PATHWAY BLACK MG | ARTIFICIAL PANCREAS DEVICE SYSTEM, SINGLE HORMONAL CONTROL | OZP | MEDTRONIC PUERTO RICO OPERATIONS CO. | MMT-1780KPK | HG40BYV | 000000763000283520 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 60 YR | Female |