FDA Adverse Event Injury Summary report: N

VERSAFITCUP CC TRIO FLAT PE HC LINER 36 / E

MDR report key: 12432603 · Received September 7, 2021

Report

Report Number
3005180920-2021-00715
Event Type
Injury
Date Received
September 7, 2021
Date of Event
August 10, 2021
Report Date
September 7, 2021
Manufacturer
MEDACTA INTERNATIONAL SA
Product Code
LZO
UDI-DI
07630030807671
PMA / PMN Number
K103352
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
US
Reporter Occupation
003

Narratives

Additional Manufacturer Narrative · 1

BATCH REVIEW PERFORMED ON 13 AUGUST 2021: LOT 1908362: (B)(4) ITEMS MANUFACTURED AND RELEASED ON 7-NOV-2019. EXPIRATION DATE: 2024-10-19. NO ANOMALIES FOUND RELATED TO THE PROBLEM. TO DATE, (B)(4) ITEMS OF THE SAME LOT HAVE BEEN ALREADY SOLD WITHOUT ANY SIMILAR REPORTED EVENT. ADDITIONAL DEVICE INVOLVED: BATCH REVIEW PERFORMED ON 13 AUGUST 2021: BALL HEADS: MECTACER 01.29.210 BIOLOX DELTA CERAMIC BALL HEAD 12/14 Ø 36 SIZE L + 4 (K112115) LOT 1905651: (B)(4) ITEMS MANUFACTURED AND RELEASED ON 15-OCT-2019. EXPIRATION DATE: 2024-10-02. NO ANOMALIES FOUND RELATED TO THE PROBLEM. TO DATE, (B)(4) ITEMS OF THE SAME LOT HAVE BEEN SOLD WITH ONE SIMILAR REPORTED EVENT.

Description of Event or Problem · 1

1 YEAR AND 1 MONTH AFTER THE PRIMARY SURGERY THE PATIENT CAME IN DUE TO SIGNS OF AN INFECTION AND THE PATHOGEN IS UNKNOWN. THE SURGEON PERFORMED A WASHOUT AND REVISED THE HEAD AND LINER. THE SURGERY WAS COMPLETED SUCCESSFULLY.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1331267 VERSAFITCUP CC TRIO FLAT PE HC LINER 36 / E HIP SHELL LINER LZO MEDACTA INTERNATIONAL SA 01.26.3644HCT 1908362 07630030807671

Patients

Seq Age Sex Outcome Treatment
1 Required Intervention