FDA Adverse Event
Malfunction
Summary report: N
SWAN-GANZ BIPOLAR PACING CATHETER FOR SVC INSERTION
MDR report key: 1243199
·
Received November 10, 2008
Report
- Report Number
- 6000002-2008-09229
- Event Type
- Malfunction
- Date Received
- November 10, 2008
- Date of Event
- October 9, 2008
- Report Date
- October 10, 2008
- Manufacturer
- EDWARDS LIFESCIENCES, PR
- Product Code
- DYG
- PMA / PMN Number
- K813521
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- JA
- Reporter Occupation
- UNKNOWN
Narratives
Additional Manufacturer Narrative · 1
THE DEVICE WAS RETURNED AND EVALUATED. CUSTOMER REPORT WAS CONFIRMED. CONTINUITY TESTING CONFIRMED THERE TO BE A SHORT CONDITION BETWEEN THE DISTAL AND PROXIMAL ELECTRODES. FURTHER TESTING CONFIRMED THE SHORT CONDITION WAS INSIDE THE Y FITTING.
Description of Event or Problem · 1
IT WAS REPORTED THAT THE CATHETER WAS UNABLE TO PACE FROM THE BEGINNING. ANOTHER DEVICE WAS USED AND THE PROBLEM WAS SOLVED. THERE WERE NO PATIENT COMPLICATIONS REPORTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | SWAN-GANZ BIPOLAR PACING CATHETER FOR SVC INSERTION | PACING CATHETER | DYG | EDWARDS LIFESCIENCES, PR | PE074F5 | 58507027 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | UNK | Other |