FDA Adverse Event Injury Summary report: N

INFINITI VISION SYSTEM

MDR report key: 1243146 · Received November 21, 2008

Report

Report Number
2028159-2008-00420
Event Type
Injury
Date Received
November 21, 2008
Date of Event
October 21, 2008
Report Date
October 23, 2008
Manufacturer
ALCON - IRVINE TECHNOLOGY CENTER
Product Code
HQC
PMA / PMN Number
K021566
Removal / Correction Number
NA
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CA
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

THE SYSTEM HAD SCHEDULED PREVENTATIVE MAINTENANCE COMPLETED THE WEEK PRIOR TO THE EVENT. THE SYSTEM WAS TESTED AND MET ALL PRODUCT SPECIFICATIONS. CORNEAL BURN IS AN ISSUE THAT IS OCCASIONALLY REPORTED WITH CATARACT SURGERY. ACCORDING TO THE ECRI HEALTH DEVICES, HAZARD UPDATE: SCLERAL AND CORNEAL BURNS DURING PHACOEMULSIFICATION, NOVEMBER 1996, VOL. 25, NO. 11: 426-431, MOST CORNEAL BURNS CAN BE TRACED TO ISSUES RELATED TO SURGICAL TECHNIQUE AND NOT TO MALFUNCTIONING EQUIPMENT. A LETTER AND COPY OF THIS INDUSTRY ARTICLE WAS PROVIDED TO THE CUSTOMER. THIS REPORT WAS MAILED TO FDA ON: 11/21/2008.

Description of Event or Problem · 1

THE SURGEON REPORTED A CORNEAL BURN OCCURRED WITHIN MICRO SECONDS OF PHACO. ADDITIONAL INFORMATION RECEIVED STATED THE CORNEAL BURN IS ON THE RIGHT SUPERIOR EDGE OF THE WOUND. NO WOUND LEAKAGE WAS DETECTED AT THE END OF SURGERY AND THE FIRST POST OPERATIVE DAY. HOWEVER, THE IOP WAS 01MMHG WITH SEIDEL NEGATIVE AND RETINAL FOLDS NOTED. THE PATIENT WAS SEEN THE SECOND POST OPERATIVE DAY WITH SEIDEL POSITIVE AND RETINAL FOLDS PRESENT. THE WOUND WAS SUTURED AND THE VISUAL ACTIVITY WAS 20/40 WITH THE IOP AT 12MMHG AND NO RETINAL FOLDS PRESENT. THE PATIENT OUTCOME AND PROGNOSIS IS GOOD.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 INFINITI VISION SYSTEM PHACOFRAGMENTATION SYSTEM HQC ALCON - IRVINE TECHNOLOGY CENTER INFINITI NA

Patients

Seq Age Sex Outcome Treatment
1 61 YR Required Intervention BSS| PRO VISC| NEVANAC| KETOROLAC