FDA Adverse Event
Malfunction
Summary report: N
RSVR MMT-103A 10PK 3CC 13L STRL
MDR report key: 1243126
·
Received November 21, 2008
Report
- Report Number
- 3004209178-2008-00979
- Event Type
- Malfunction
- Date Received
- November 21, 2008
- Date of Event
- November 6, 2008
- Report Date
- November 6, 2008
- Manufacturer
- MEDTRONIC PUERTO RICO OPERATIONS MED-REL
- Product Code
- FMF
- PMA / PMN Number
- K991936
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Occupation
- NOT APPLICABLE
Narratives
Additional Manufacturer Narrative · 1
A COMPLETE ANALYSIS AND TESTING OF THREE SEALED RESERVOIRS SHOWED THAT THEY WERE FUNCTIONING PROPERLY AND PASSED ALL FUNCTIONAL TESTING. AFTER TESTING, IT WAS CONCLUDED THAT THE RESERVOIRS OPERATED WITHIN SPECIFICATIONS. NO LEAKING WAS OBSERVED DURING ANALYSIS. HOWEVER, AN ANALYSIS OF A USED RESERVOIR REVEALED THAT THE RESERVOIR HAD LEAKED DUE TO AN UNK STICKY SUBSTANCE FOUND ON BOTH O-RINGS AND PLUNGER ROD.
Description of Event or Problem · 1
THE CUSTOMER STATED THAT INSULIN WAS LEAKING AROUND THE PLUNGER OF THE RESERVOIR. NOTHING FURTHER WAS REPORTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | RSVR MMT-103A 10PK 3CC 13L STRL | RESERVOIR | FMF | MEDTRONIC PUERTO RICO OPERATIONS MED-REL | MMT-103A | H7201022 |