FDA Adverse Event Malfunction Summary report: N

840 VENTILATOR

MDR report key: 1242874 · Received November 19, 2008

Report

Report Number
8020893-2008-00612
Event Type
Malfunction
Date Received
November 19, 2008
Date of Event
November 10, 2008
Report Date
November 10, 2008
Manufacturer
PURITAN-BENNETT CORP.
Product Code
CBK
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
SC, US
Reporter Occupation
OTHER

Narratives

Description of Event or Problem · 1

THE SERVICE REPORT SHOWS THE CUSTOMER REPORTED THAT THE 840 VENTILATOR STOPPED CYCLING WHILE IN USE ON A PATIENT. THE PATIENT WAS NOT HARMED OR INJURED AS A RESULT OF THE EVENT. THE NELLCOR PURITAN BENNETT CUSTOMER SUPPORT ENGINEER (CSE) VERIFIED THE MALFUNCTION. THE CSE INSPECTED THE DEVICE AND REPLACED THE BDU CPU PCB. THE UNIT PASSED EXTENDED SELF TESTING.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 840 VENTILATOR VENTILATOR CBK PURITAN-BENNETT CORP. 840

Patients

Seq Age Sex Outcome Treatment
1