BD PLASTIPAK¿ 20ML SYRINGE LUER-LOK¿
Report
- Report Number
- 3003152976-2021-00564
- Event Type
- Malfunction
- Date Received
- September 4, 2021
- Date of Event
- July 8, 2021
- Report Date
- October 13, 2021
- Manufacturer
- BECTON DICKINSON, S.A.
- Product Code
- FMF
- PMA / PMN Number
- NA
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- FR
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
- Health Professional
- Yes
Narratives
H.6. INVESTIGATION: NO SAMPLES OR PHOTOS RECEIVED FOR INVESTIGATION. A DEVICE HISTORY REVIEW WAS PERFORMED FOR THE REPORTED LOT 2102073 NO DEVIATIONS OR NON-CONFORMANCES WERE IDENTIFIED DURING THE MANUFACTURING PROCESS THAT COULD HAVE CONTRIBUTED TO THIS ISSUE. DURING MANUFACTURING PROCESS BREAK OUT FORCE, SUSTAINING FORCE AND SILICONE CONTENT TESTS ARE PERFORMED, RESULTS FOR LOT 2103048 ARE WITHIN SPECIFICATION LIMITS. FINAL PRODUCTS IN THIS MANUFACTURING LINE, FOR THIS REFERENCE ARE SAMPLED AND THEY ARE SUBJECTED TO VISUAL AND FUNCTIONAL INSPECTIONS DURING THE DIFFERENT MANUFACTURING SUB-PROCESSES ACCORDING TO PROCEDURES. BASED ON THE INVESTIGATION RESULTS, NO MANUFACTURING RELATED DEFECTS COULD BE IDENTIFIED AND THEREFORE, A CAUSE FOR THE REPORTED INCIDENT COULD NOT BE DETERMINED.
IT WAS REPORTED THAT BD PLASTIPAK¿ 20ML SYRINGE LUER-LOK¿ HAD A DIFFICULT TO MOVE PLUNGER. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER: "WHEN PUSHING THE PLUNGER TO READJUST THE VOLUME AFTER TAKING THE CHEMOTHERAPY, IT GETS STUCK".
A DEVICE EVALUATION AND/OR DEVICE HISTORY REVIEW IS ANTICIPATED, BUT IS NOT COMPLETE. UPON COMPLETION, A SUPPLEMENTAL REPORT WILL BE FILED.
IT WAS REPORTED THAT BD PLASTIPAK¿ 20ML SYRINGE LUER-LOK¿ HAD A DIFFICULT TO MOVE PLUNGER. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER: "WHEN PUSHING THE PLUNGER TO READJUST THE VOLUME AFTER TAKING THE CHEMOTHERAPY, IT GETS STUCK. "
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1323546 | BD PLASTIPAK¿ 20ML SYRINGE LUER-LOK¿ | PISTON SYRINGE | FMF | BECTON DICKINSON, S.A. | 2102073 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Unknown |