FDA Adverse Event Malfunction Summary report: N

BD PLASTIPAK¿ 20ML SYRINGE LUER-LOK¿

MDR report key: 12426322 · Received September 4, 2021

Report

Report Number
3003152976-2021-00564
Event Type
Malfunction
Date Received
September 4, 2021
Date of Event
July 8, 2021
Report Date
October 13, 2021
Manufacturer
BECTON DICKINSON, S.A.
Product Code
FMF
PMA / PMN Number
NA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
FR
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

H.6. INVESTIGATION: NO SAMPLES OR PHOTOS RECEIVED FOR INVESTIGATION. A DEVICE HISTORY REVIEW WAS PERFORMED FOR THE REPORTED LOT 2102073 NO DEVIATIONS OR NON-CONFORMANCES WERE IDENTIFIED DURING THE MANUFACTURING PROCESS THAT COULD HAVE CONTRIBUTED TO THIS ISSUE. DURING MANUFACTURING PROCESS BREAK OUT FORCE, SUSTAINING FORCE AND SILICONE CONTENT TESTS ARE PERFORMED, RESULTS FOR LOT 2103048 ARE WITHIN SPECIFICATION LIMITS. FINAL PRODUCTS IN THIS MANUFACTURING LINE, FOR THIS REFERENCE ARE SAMPLED AND THEY ARE SUBJECTED TO VISUAL AND FUNCTIONAL INSPECTIONS DURING THE DIFFERENT MANUFACTURING SUB-PROCESSES ACCORDING TO PROCEDURES. BASED ON THE INVESTIGATION RESULTS, NO MANUFACTURING RELATED DEFECTS COULD BE IDENTIFIED AND THEREFORE, A CAUSE FOR THE REPORTED INCIDENT COULD NOT BE DETERMINED.

Description of Event or Problem · 0

IT WAS REPORTED THAT BD PLASTIPAK¿ 20ML SYRINGE LUER-LOK¿ HAD A DIFFICULT TO MOVE PLUNGER. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER: "WHEN PUSHING THE PLUNGER TO READJUST THE VOLUME AFTER TAKING THE CHEMOTHERAPY, IT GETS STUCK".

Additional Manufacturer Narrative · 1

A DEVICE EVALUATION AND/OR DEVICE HISTORY REVIEW IS ANTICIPATED, BUT IS NOT COMPLETE. UPON COMPLETION, A SUPPLEMENTAL REPORT WILL BE FILED.

Description of Event or Problem · 1

IT WAS REPORTED THAT BD PLASTIPAK¿ 20ML SYRINGE LUER-LOK¿ HAD A DIFFICULT TO MOVE PLUNGER. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER: "WHEN PUSHING THE PLUNGER TO READJUST THE VOLUME AFTER TAKING THE CHEMOTHERAPY, IT GETS STUCK. "

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1323546 BD PLASTIPAK¿ 20ML SYRINGE LUER-LOK¿ PISTON SYRINGE FMF BECTON DICKINSON, S.A. 2102073

Patients

Seq Age Sex Outcome Treatment
1 Unknown