25G BI-B POSTERIOR ELITE W WF
Report
- Report Number
- 0001920664-2021-00107
- Event Type
- Injury
- Date Received
- September 3, 2021
- Report Date
- September 3, 2021
- Manufacturer
- BAUSCH + LOMB, INC.
- Product Code
- HQC
- PMA / PMN Number
- K170052
- Adverse Event
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- FR
- Reporter Occupation
- OTHER
Narratives
THE DEVICE WAS NOT RETURNED FOR EVALUATION, SO UNABLE TO INVESTIGATE FOR ROOT CAUSE. THE REVIEW OF THE DEVICE HISTORY RECORDS (DHR) FOR ASSEMBLY OF THE FINISHED SE5425WB LOT W8735 SHOWED THE PRODUCT WAS PROPERLY MANUFACTURED, TESTED, AND MET ALL SPECIFICATIONS AT THE TIME OF RELEASE. THE REVIEW OF THE DEVICE HISTORY RECORDS FOR THE MANUFACTURE OF THE LIGHT PIPE PART # (B)(4)SHOWED THAT FOR EACH OF THE SIX LOTS MANUFACTURED AND USED IN LOT W8735 (LIGHT PIPE LOTS M0034-152, -203, -204, -345, -462, AND -463) THAT THE RECORD WERE ACCEPTABLE, ALL FUNCTIONAL TESTING AND VISUAL INSPECTION WAS PERFORMED PER PROCEDURE, NO NEGATIVE QUALITY PATTERNS WERE REVIEWED AND NO NON-CONFORMANCES ARE ASSOCIATED WITH THE DEVICE HISTORY RECORDS. THE LOT HISTORY, TREND ANALYSIS, RISK ANALYSIS AND/OR DIRECTIONS FOR USE REVIEW WERE CONSIDERED ACCEPTABLE, WITH THE PRODUCT PERFORMING WITHIN ANTICIPATED RATES. NO CORRECTIVE ACTION REQUIRED. (REPORT 3 OF 4 0001920664-2021-00107) SEE RELATED REPORTS 0001920664-2021-00105, 0001920664-2021-00106, 0001920664-2021-00108.
THE USER FACILITY REPORTED THE TIP OF THE LIGHT PIPE BROKE AFTER A FEW SECONDS OF USE. THE TIP DID NOT BREAK IN THE PATIENTS' EYE AND NO SUTURES WERE REQUIRED. THE SURGERY PROCEEDED WITH A DIFFERENT PACK. THE SURGERY WAS EXTENDED BY 10 MINUTES AND THE ANESTHESIA WAS PROLONGED. ASIDE FROM THE DELAY IN SURGERY, THERE WAS NO IMPACT TO THE PATIENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1319538 | 25G BI-B POSTERIOR ELITE W WF | UNIT, PHACOFRAGMENTATION | HQC | BAUSCH + LOMB, INC. | SE5425WVB | 222 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Required Intervention |