NXSTAGE SYSTEM ONE
Report
- Report Number
- 3003464075-2008-00600
- Event Type
- Injury
- Date Received
- November 21, 2008
- Date of Event
- October 23, 2008
- Report Date
- October 23, 2008
- Manufacturer
- NXSTAGE MEDICAL, INC.
- Product Code
- KDI
- PMA / PMN Number
- K050525
- Adverse Event
- Yes
- Report Source
- Manufacturer report
- Reporter Occupation
- NOT APPLICABLE
Narratives
THE REPORTED BLOOD LOSS IS ATTRIBUTED TO THE OPERATOR NOT COMPLETING RINSEBACK OF THE PATIENT'S BLOOD AS INSTRUCTED IN THE USER'S GUIDE. PRODUCT WAS NOT RETURNED AT THE TIME OF THIS REPORT FOR EVALUATION AS REQUESTED. THE REPORTED LEAK CANNOT BE CONFIRMED. THE USER'S GUIDE INCLUDES ADEQUATE WARNINGS FOR THE OPERATOR TO PERIODICALLY CHECK THE SYSTEM FOR LEAKS AS THE CYCLER MAY NOT SENSE SLOW LEAKS AND TO TERMINATE THE TREATMENT IF A LEAK CANNOT BE RESOLVED. A DIRECT CORRELATION BETWEEN NXSTAGE SYSTEM ONE AND THE REPORTED BLOOD LOSS CANNOT BE ESTABLISHED. FACILITY STAFF HAS BEEN NOTIFIED. NXSTAGE MEDICAL CONSIDERS THIS REPORT CLOSED. NO ADDITIONAL INFORMATION WILL BE PROVIDED.
THIS REPORT IS BEING FILED AS AN ADVERSE EVENT SOLELY TO COMPLY WITH THE FDA'S BLOOD LOSS POLICY. DURING A ROUTINE HEMODIALYSIS TREATMENT, THE CYCLER POWERED OFF. A CLEAR FLUID LEAK WAS OBSERVED. TREATMENT WAS DISCONTINUED WITHOUT PERFORMING RINSEBACK, DUE TO POSSIBLE CLOTTING RESULTING IN AN ESTIMATED BLOOD LOSS OF 190CC. NO MEDICAL INTERVENTION WAS REQUIRED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | NXSTAGE SYSTEM ONE | HIGH PERMEABILITY HEMODIALYSIS SYSTEM | KDI | NXSTAGE MEDICAL, INC. | CAR-170 | 8087701 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 47 YR | Other |