FDA Adverse Event Injury Summary report: N

STYLE 115 SILICONE GEL FILLED BREAST IMPLANT

MDR report key: 12421608 · Received September 3, 2021

Report

Report Number
9617229-2021-49673
Event Type
Injury
Date Received
September 3, 2021
Date of Event
June 9, 2021
Report Date
October 7, 2021
Manufacturer
ALLERGAN (COSTA RICA)
Product Code
FTR
PMA / PMN Number
P020056
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
MN, US
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 0

DEVICE EVALUATION: VISUAL ANALYSIS OF THE RETURNED DEVICE IDENTIFIED ONE EXTENDED OPENING, FOLD CREASES AND BROWN PARTICLES MATERIAL WAS FOUND ON THE SHELL. A WEIGHT TEST OF THE DEVICE WAS VERIFIED AND THE DEVICE UNDERWEIGHT. A MICROSCOPIC ANALYSIS WAS PERFORMED WHICH IDENTIFY ONE SHARP EDGE OPENING IN THE SHELL WITH STRESS MARKS. A DIMENSION MEASUREMENT IN THE SHELL WAS PERFORMED WHICH IDENTIFY THE THICKNESS WITHIN SPECIFICATION. BASED ON THE DEVICE ANALYSIS THE FINAL ASSESSMENT IS A SHARP EDGE OPENING IN THE SHELL WITH STRESS MARKS IN THE SIDE POSTERIOR ASSESSED AS SURGICAL IMPACT OPENING.

Description of Event or Problem · 0

HEALTHCARE PROFESSIONAL REPORTED LEFT SIDE EXCHANGE FROM TEXTURED TO SMOOTH DUE TO PATIENT CONCERN WITH THE PRODUCT. HEALTHCARE PROFESSIONAL REPORTED ADDITIONAL EVENT OF RUPTURE. DEVICE HAS BEEN EXPLANTED.

Additional Manufacturer Narrative · 1

A REVIEW OF THE DEVICE HISTORY RECORD HAS BEEN COMPLETED. NO DEVIATIONS OR NON-CONFORMANCES NOTED. FURTHER INFORMATION FROM THE REPORTER REGARDING EVENT, PRODUCT, OR PATIENT DETAILS HAS BEEN REQUESTED. NO ADDITIONAL INFORMATION IS AVAILABLE AT THIS TIME. REASON FOR REOPERATION: PATIENTS CONCERN WITH PRODUCT AND RUPTURE.

Description of Event or Problem · 1

HEALTHCARE PROFESSIONAL REPORTED LEFT SIDE EXCHANGE FROM TEXTURED TO SMOOTH DUE TO PATIENT CONCERN WITH THE PRODUCT. HEALTHCARE PROFESSIONAL REPORTED ADDITIONAL EVENT OF RUPTURE. DEVICE HAS BEEN EXPLANTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1320571 STYLE 115 SILICONE GEL FILLED BREAST IMPLANT PROSTHESIS, BREAST, NONINFLATABLE, INTERNAL, SILICONE GEL-FILLED FTR ALLERGAN (COSTA RICA) 2277728

Patients

Seq Age Sex Outcome Treatment
1 40 YR Required Intervention