FDA Adverse Event Malfunction Summary report: N

ENDOEYE FLEX 3D DEFLECTABLE VIDEOSCOPE

MDR report key: 12419343 · Received September 3, 2021

Report

Report Number
8010047-2021-11195
Event Type
Malfunction
Date Received
September 3, 2021
Date of Event
August 3, 2021
Report Date
November 16, 2021
Manufacturer
OLYMPUS MEDICAL SYSTEMS CORP.
Product Code
GCJ
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
HK
Reporter Occupation
OTHER
Health Professional
N

Narratives

Additional Manufacturer Narrative · 0

THIS SUPPLEMENTAL REPORT IS BEING SUBMITTED TO PROVIDE ADDITIONAL INFORMATION. THE SUBJECT DEVICE HAS NOT BEEN RETURNED TO OLYMPUS MEDICAL SYSTEMS CORP. (OMSC) BUT WAS RETURNED TO OLYMPUS HONG KONG AND CHINA (OHC). OMSC REVIEWED THE MANUFACTURING HISTORY (DHR) OF THE SUBJECT DEVICE AND CONFIRMED NO IRREGULARITY. THEREFORE, THE EXACT CAUSE OF THE REPORTED PHENOMENON COULD NOT BE CONCLUSIVELY DETERMINED. HOWEVER, BASED UPON THE INFORMATION FROM OHC AND SIMILAR PAST CASES, OMSC SURMISED THAT THIS PHENOMENON OCCURRED BECAUSE THE CABLE WAS DAMAGED AND THE IMAGE SIGNAL WAS NOT PROCESSED PROPERLY. IF ADDITIONAL INFORMATION IS RECEIVED, THIS REPORT WILL BE SUPPLEMENTED.

Additional Manufacturer Narrative · 1

THE SUBJECT DEVICE HAS NOT BEEN RETURNED TO OMSC FOR EVALUATION BUT WAS RETURNED TO OLYMPUS (B)(4) AND (B)(4) (OHC). OHC CHECKED THE SUBJECT DEVICE AND FOUND THAT THE REPORTED PHENOMENON WAS DUPLICATED DUE TO THE DAMAGE OF THE CCD UNIT/CABLE. THE EXACT CAUSE HAS BEEN UNDER INVESTIGATION. THEREFORE, THE EXACT CAUSE OF THE REPORTED EVENT COULD NOT BE CONCLUSIVELY DETERMINED AT THIS TIME. IF ADDITIONAL INFORMATION IS RECEIVED, THIS REPORT WILL BE SUPPLEMENTED.

Description of Event or Problem · 1

OLYMPUS MEDICAL SYSTEMS CORP. (OMSC) WAS INFORMED FROM THE USER FACILITY THAT DURING THE INSPECTION BEFORE THE USE, IT WAS FOUND THAT THE SCOPE COMMUNICATION ERROR E226 OCCURRED ON THE SUBJECT DEVICE. THERE WAS NO REPORT OF PATIENT INJURY ASSOCIATED WITH THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1319604 ENDOEYE FLEX 3D DEFLECTABLE VIDEOSCOPE FLEX DEFLECTABLE VIDEOSCOPE GCJ OLYMPUS MEDICAL SYSTEMS CORP. LTF-S300-10-3D

Patients

Seq Age Sex Outcome Treatment
1 Unknown