CLINICAL INFORMATION CENTER
Report
- Report Number
- 2124823-2008-00093
- Event Type
- Malfunction
- Date Received
- November 19, 2008
- Date of Event
- October 26, 2008
- Report Date
- November 19, 2008
- Manufacturer
- GE HEALTHCARE
- Product Code
- DSI
- PMA / PMN Number
- K062976
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- FL, US
- Reporter Occupation
- BIOMEDICAL ENGINEER
Narratives
IT WAS REPORTED THAT TIME WOULD NOT ADVANCE PAST 02:00 ON THE CIC. THE USERS NOTED PT WAVEFORMS WERE DROPPING OUT ON THE CIC FOR BOTH HARDWIRE AND TELEMETRY MONITORING. THE WAVEFORMS WOULD ALLEGEDLY DISAPPEAR FOR APPROXIMATELY ONE MINUTE, AND THEN WOULD REAPPEAR. THE REPORTED EVENT RESEMBLES A KNOWN CONDITION WHERE THE CIC TIME MASTER ON THE UNITY NETWORK BROADCASTS TIME CHANGES AND THE CIC REFLECTS THESE CHANGES, CAUSING MULTIPLE TIME REQUESTS AND EXCESSIVE NETWORK TRAFFIC. THE CONDITION CAN LEAD TO A LOSS OF MONITORING ON THE UNITY NETWORK, WHICH MAY DELAY RECOGNITION AND TREATMENT OF A CRITICAL PT EVENT. NO PT INJURY WAS REPORTED. GE HEALTHCARE'S INVESTIGATION INTO THE REPORTED OCCURRENCE IS STILL ONGOING. A FOLLOW-UP REPORT WILL BE ISSUED WHEN THE INVESTIGATION HAS BEEN COMPLETED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | CLINICAL INFORMATION CENTER | CENTRAL MONITORING SYSTEM | DSI | GE HEALTHCARE | 2019989-003 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |