FDA Adverse Event Malfunction Summary report: N

CLINICAL INFORMATION CENTER

MDR report key: 1241915 · Received November 19, 2008

Report

Report Number
2124823-2008-00093
Event Type
Malfunction
Date Received
November 19, 2008
Date of Event
October 26, 2008
Report Date
November 19, 2008
Manufacturer
GE HEALTHCARE
Product Code
DSI
PMA / PMN Number
K062976
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
FL, US
Reporter Occupation
BIOMEDICAL ENGINEER

Narratives

Description of Event or Problem · 1

IT WAS REPORTED THAT TIME WOULD NOT ADVANCE PAST 02:00 ON THE CIC. THE USERS NOTED PT WAVEFORMS WERE DROPPING OUT ON THE CIC FOR BOTH HARDWIRE AND TELEMETRY MONITORING. THE WAVEFORMS WOULD ALLEGEDLY DISAPPEAR FOR APPROXIMATELY ONE MINUTE, AND THEN WOULD REAPPEAR. THE REPORTED EVENT RESEMBLES A KNOWN CONDITION WHERE THE CIC TIME MASTER ON THE UNITY NETWORK BROADCASTS TIME CHANGES AND THE CIC REFLECTS THESE CHANGES, CAUSING MULTIPLE TIME REQUESTS AND EXCESSIVE NETWORK TRAFFIC. THE CONDITION CAN LEAD TO A LOSS OF MONITORING ON THE UNITY NETWORK, WHICH MAY DELAY RECOGNITION AND TREATMENT OF A CRITICAL PT EVENT. NO PT INJURY WAS REPORTED. GE HEALTHCARE'S INVESTIGATION INTO THE REPORTED OCCURRENCE IS STILL ONGOING. A FOLLOW-UP REPORT WILL BE ISSUED WHEN THE INVESTIGATION HAS BEEN COMPLETED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 CLINICAL INFORMATION CENTER CENTRAL MONITORING SYSTEM DSI GE HEALTHCARE 2019989-003

Patients

Seq Age Sex Outcome Treatment
1