LAP-BAND AP ADJUSTABLE GASTRIC BANDING SYSTEM (SMALL) ACCESS PORT I
Report
- Report Number
- 2024601-2008-00911
- Event Type
- Malfunction
- Date Received
- November 18, 2008
- Date of Event
- October 2, 2008
- Report Date
- October 22, 2008
- Manufacturer
- COSTA RICA
- Product Code
- LTI
- PMA / PMN Number
- P000008
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- FL, US
- Reporter Occupation
- OTHER
Narratives
TAPER II. VISUAL EXAMINATION OF THE RETURNED DEVICE DETERMINED THE ACCESS PORT TUBING CONNECTOR TO BE A TAPER II. ANALYSIS OF THE DEVICE NOTED DAMAGED FROM A SHARP INSTRUMENT TO THE BAND TUBING. ALSO NOTED WERE NON-PENETRATING NICKS TO THE PORT SEPTUM. THE DAMAGE APPEARS TO BE CAUSED BY A NEEDLE. THERE WAS NO INDICATION OF WEAR RELATED DAMAGE TO THE PORTION OF THE LAP-BAND SYSTEM RETURNED. PERFORMANCE TESTS INDICATE NO LEAKAGE AT THE ACCESS PORT OR ACCESS PORT TUBING. THE LAB WAS UNABLE TO DUPLICATE OR CONFIRM THE REPORTED EVENT. THERE IS NO OTHER INFORMATION AVAILABLE AT THIS TIME FROM THE SURGEON TO DETERMINE THE CAUSE OF THE ALLEGED EVENT. DEVICE LABELING ADDRESSES THE POSSIBLE OUTCOME OF LEAKAGE AS FOLLOWS: "DEFLATION OF THE BAND MAY OCCUR DUE TO LEAKAGE FROM THE BAND, THE PORT, OR THE CONNECTOR TUBING". IN ADDITION, THE LABELING ALSO ADDRESSES THE FOLLOWING POSSIBLE OUTCOME OF ACCESS PORT LEAKAGE: "CAUTION: USE OF AN INAPPROPRIATE NEEDLE MAY CAUSE ACCESS PORT LEAKAGE AND REQUIRE REOPERATION TO REPLACE THE PORT. DO NOT USE STANDARD HYPODERMIC NEEDLES AS THESE MAY CAUSE LEAKS. USE ONLY BIOENTERICS LAP-BAND SYSTEM ACCESS PORT NEEDLES."
REPORTED AS A LEAKING PORT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | LAP-BAND AP ADJUSTABLE GASTRIC BANDING SYSTEM (SMALL) ACCESS PORT I | DEVICE FOR TREATMENT OF MORBID OBESITY | LTI | COSTA RICA | NA | 1449173 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 48 YR |