COLLEAGUE VOLUMETRIC INFUSION PUMP
Report
- Report Number
- 6000001-2007-04387
- Event Type
- Malfunction
- Date Received
- November 25, 2008
- Date of Event
- January 31, 2007
- Report Date
- January 31, 2007
- Manufacturer
- BAXTER HEALTHCARE PTE.LTD.
- Product Code
- FRN
- Removal / Correction Number
- 6000001-7/20/05-015-C
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- FL, US
- Reporter Occupation
- OTHER
Narratives
EVALUATION SUMMARY: DURING PRODUCT EVALUATION, A DEFECTIVE USER INTERFACE MODULE PRINTED CIRCUIT BOARD (UIM PCB) WAS OBSERVED. FAILURE CODE 703 IN THE EVENT HISTORY CONFIRMS THE DEFECTIVE UIM PCB. THIS FAILURE CODE IS MANIFESTED AS A RESULT OF THE UIM PCB BEING DEFECTIVE. THE UIM PCB WAS REPLACED. REVIEW OF THE COMPLAINT HISTORY REVEALS SIMILAR REPORTS HAVE BEEN RECEIVED FOR THIS PRODUCT FOR THE REPORTED ISSUE. THIS ISSUE IS BEING INVESTIGATED UNDER CAPA.
THE DEVICE WAS RETURNED TO BAXTER FOR SERVICE. DURING TESTING, THE BAXTER TECHNICIAN REPORTED AN INFUSION PUMP WITH A DEFECTIVE USER INTERFACE MECHANISM PRINTED CIRCUIT BOARD (UIM PCB). ACCORDING TO THE HOSPITAL REPRESENTATIVE, NO PATIENT INJURY OR MEDICAL INTERVENTION HAS BEEN REPORTED RELATED TO THIS DEVICE. NO ADDITIONAL INFORMATION IS KNOWN.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | COLLEAGUE VOLUMETRIC INFUSION PUMP | INFUSION PUMP | FRN | BAXTER HEALTHCARE PTE.LTD. | N/A | NA |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |