FDA Adverse Event
Malfunction
Summary report: N
COLLEAGUE 3 VOLUMETRIC INFUSION PUMP
MDR report key: 1241176
·
Received November 25, 2008
Report
- Report Number
- 6000001-2007-04210
- Event Type
- Malfunction
- Date Received
- November 25, 2008
- Date of Event
- January 23, 2007
- Report Date
- January 23, 2007
- Manufacturer
- BAXTER HEALTHCARE PTE.LTD.
- Product Code
- FRN
- Removal / Correction Number
- 6000001-7/20/05-015-C
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- CA, US
- Reporter Occupation
- OTHER
Narratives
Additional Manufacturer Narrative · 1
EVALUATION SUMMARY: EVALUATION WAS PERFORMED. INSPECTION OF THE PUMP REVEALED DEFECTIVE USER INTERFACE MODULE PRINTER CIRCUIT BOARD (UIM PCB) WAS DUE TO FAILURE CODE 703:00. THE UIM PCB WAS REPLACED. THE PUMP WAS TESTED FOR PROPER OPERATION AND PUMP FOUND TO FUNCTION PROPERLY. REVIEW OF THE COMPLAINT HISTORY REVEALS SIMILAR REPORTS HAVE BEEN RECEIVED FOR THIS PRODUCT FOR THE REPORTED ISSUE. THIS ISSUE IS CURRENTLY BEING INVESTIGATED.
Description of Event or Problem · 1
DURING SERVICE BY BAXTER, DEFECTIVE USER INTERFACE MODULE PRINTER CIRCUIT BOARD (UIM PCB) WAS FOUND. ACCORDING TO THE HOSPITAL REPRESENTATIVE, NO PATIENT INJURY OR MEDICAL INTERVENTION HAD BEEN REPORTED RELATED TO THE DEVICE. NO ADDITIONAL INFORMATION OR CONTACT WAS AVAILABLE.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | COLLEAGUE 3 VOLUMETRIC INFUSION PUMP | INFUSION PUMP | FRN | BAXTER HEALTHCARE PTE.LTD. | N/A | NA |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |