FDA Adverse Event
Injury
Summary report: N
PUMP MMT-722NAB PRDGM INS V2.2 BL EN
MDR report key: 1241058
·
Received November 21, 2008
Report
- Report Number
- 3004209178-2008-00987
- Event Type
- Injury
- Date Received
- November 21, 2008
- Date of Event
- October 29, 2008
- Report Date
- November 8, 2008
- Manufacturer
- MEDTRONIC PUERTO RICO OPERATIONS MED-REL
- Product Code
- LZG
- PMA / PMN Number
- P980022
- Adverse Event
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- FL, US
- Reporter Occupation
- OTHER
Narratives
Additional Manufacturer Narrative · 1
CURRENTLY, IT IS UNK WHETHER OR NOT THE DEVICE MAY HAVE CAUSED OR CONTRIBUTED TO THE EVENT AS NO PROD HAS BEEN RETURNED. NO CONCLUSION CAN BE DRAWN AT THIS TIME. WE THEREFORE CONSIDER THIS REPORT COMPLETE TO THE BEST OF OUR KNOWLEDGE.
Description of Event or Problem · 1
THE CUSTOMER STATED THAT SHE WAS HOSPITALIZED DUE TO HYPOGLYCEMIA. TROUBLESHOOTING WAS PERFORMED AND FOUND THAT THE INSULIN PUMP WAS READING THE RESERVOIR VOLUME CORRECTLY. THE PRIME TEST PASSED. THE INSULIN PUMP ALARMED NO DELIVERY FOLLOWING THE EVENT. NOTHING FURTHER WAS REPORTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | PUMP MMT-722NAB PRDGM INS V2.2 BL EN | INFUSION PUMP | LZG | MEDTRONIC PUERTO RICO OPERATIONS MED-REL | MMT-722NAB |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 76 YR | Hospitalization |