FDA Adverse Event Injury Summary report: N

IMP,TSV,3.7,8,MTX,MG

MDR report key: 12409063 · Received September 1, 2021

Report

Report Number
0002023141-2021-02403
Event Type
Injury
Date Received
September 1, 2021
Date of Event
July 16, 2021
Report Date
February 8, 2022
Manufacturer
ZIMMER DENTAL
Product Code
DZE
UDI-DI
00889024019966
PMA / PMN Number
K101977
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
IN
Reporter Occupation
DENTIST
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

SIMILAR COMPLAINTS FOR THE INABILITY TO ACHIEVE PRIMARY STABILITY (UNABLE TO PLACE IMPLANT INTO OSTEOTOMY) HAVE BEEN PREVIOUSLY INVESTIGATED. REFER TO ATTACHED SUMMARY INVESTIGATION. VISUAL AND DIMENSIONAL EVALUATIONS OF THE PREVIOUSLY RETURNED PRODUCT HAVE NOT IDENTIFIED OR SUGGESTED MANUFACTURING NON-CONFORMANCES. WHILE NON-CONFORMANCES WERE IDENTIFIED FOR SOME LOTS DURING MANUFACTURING RECORDS REVIEWS, THE DOCUMENTED DISPOSITION ACTIONS FOR EACH HAVE NOT SUGGESTED THE LIKELY RELEASE OF NON-CONFORMING PRODUCT. TO DATE, ALL COMPLAINT DATA WAS FOUND TO BE CONFORMING AND DID NOT MEET CAPA/HHE/D/IE ESCALATION. THEREFORE, THERE WERE NO COMPLAINTS WHICH CONFIRMED A MANUFACTURING OR DESIGN RELATED ISSUE CAUSED OR CONTRIBUTED TO THE REPORTED EVENT. DHR REVIEW WAS COMPLETED FOR THE SUBJECT LOT NUMBER (1231195). IT WAS CONFIRMED THAT ALL OPERATIONS AND INSPECTIONS WERE EXECUTED AS PER APPLICABLE PROCEDURE. NO DEVIATIONS OR NON-CONFORMANCES, WHICH COULD HAVE CAUSED OR CONTRIBUTED TO THE REPORTED EVENT WERE NOTED AS PART OF THE DHR. LOT WAS INSPECTED AND PASSED ALL ACCEPTANCE CRITERIA BY QA. COMPLAINT HISTORY REVIEW WAS PERFORMED FOR THE REPORTED LOT NUMBER (1231195) FOR SIMILAR EVENTS AND 2 OTHER COMPLAINTS WERE IDENTIFIED. . AS DOCUMENTED IN THE SUMMARY INVESTIGATION, CONTRIBUTING FACTORS FOR THE REPORTED EVENT LIKELY EXIST OUTSIDE OF ZIMMER BIOMET CONTROL, INCLUDING THOSE RELATED TO PATIENT BIOLOGICAL FACTORS/CONDITION AND SURGICAL TECHNIQUE. THE FOLLOWING SECTIONS HAVE BEEN UPDATED: G6: CHECKED "FOLLOW-UP". H3: CHANGED "YES" TO "NO".

Additional Manufacturer Narrative · 0

SIMILAR COMPLAINTS FOR THE INABILITY TO ACHIEVE PRIMARY STABILITY (UNABLE TO PLACE IMPLANT INTO OSTEOTOMY) HAVE BEEN PREVIOUSLY INVESTIGATED. REFER TO ATTACHED SUMMARY INVESTIGATION. VISUAL AND DIMENSIONAL EVALUATIONS OF THE PREVIOUSLY RETURNED PRODUCT HAVE NOT IDENTIFIED OR SUGGESTED MANUFACTURING NON-CONFORMANCES. WHILE NON-CONFORMANCES WERE IDENTIFIED FOR SOME LOTS DURING MANUFACTURING RECORDS REVIEWS, THE DOCUMENTED DISPOSITION ACTIONS FOR EACH HAVE NOT SUGGESTED THE LIKELY RELEASE OF NON-CONFORMING PRODUCT. TO DATE, ALL COMPLAINT DATA WAS FOUND TO BE CONFORMING AND DID NOT MEET CAPA/HHE/D/IE ESCALATION. THEREFORE, THERE WERE NO COMPLAINTS WHICH CONFIRMED A MANUFACTURING OR DESIGN RELATED ISSUE CAUSED OR CONTRIBUTED TO THE REPORTED EVENT. DHR REVIEW WAS COMPLETED FOR THE SUBJECT LOT NUMBER (1231195). IT WAS CONFIRMED THAT ALL OPERATIONS AND INSPECTIONS WERE EXECUTED AS PER APPLICABLE PROCEDURE. NO DEVIATIONS OR NON-CONFORMANCES, WHICH COULD HAVE CAUSED OR CONTRIBUTED TO THE REPORTED EVENT WERE NOTED AS PART OF THE DHR. LOT WAS INSPECTED AND PASSED ALL ACCEPTANCE CRITERIA BY QA. COMPLAINT HISTORY REVIEW WAS PERFORMED FOR THE REPORTED LOT NUMBER (1231195) FOR SIMILAR EVENTS AND 2 OTHER COMPLAINTS WERE IDENTIFIED UNDER (B)(4). AS DOCUMENTED IN THE SUMMARY INVESTIGATION, CONTRIBUTING FACTORS FOR THE REPORTED EVENT LIKELY EXIST OUTSIDE OF ZIMMER BIOMET CONTROL, INCLUDING THOSE RELATED TO PATIENT BIOLOGICAL FACTORS/CONDITION AND SURGICAL TECHNIQUE.

Description of Event or Problem · 0

NO FURTHER EVENT INFORMATION AVAILABLE AT THE TIME OF THIS REPORT.

Description of Event or Problem · 0

NO FURTHER EVENT INFORMATION AVAILABLE AT THE TIME OF THIS REPORT.

Additional Manufacturer Narrative · 1

ZIMMER BIOMET COMPLAINT NUMBER (B)(4). A SUMMARY INVESTIGATION HAS BEEN COMPLETED FOR LACK OF PRIMARY STABILITY EVENTS THAT DO NOT ALLEGE A DEFICIENCY WITH THE IMPLANT AND IDENTIFIED THAT THE REPORTED EVENT IS LIKELY DUE TO BIOLOGICAL FACTORS WHICH HAVE AN ADVERSE EFFECT ON IMPLANT STABILITY OR IS RELATED TO SURGICAL TECHNIQUE AND INSERTION TORQUE. DUE TO A WIDE RANGE OF EXTERNAL (NON-DESIGN/ NON-MANUFACTURING RELATED), IDENTIFYING A DEFINITIVE ROOT CAUSE IS GENERALLY NOT POSSIBLE. SHOULD ADDITIONAL INFORMATION BE RECEIVED WHICH INDICATES THAT THE DEVICE MAY HAVE CAUSED OR CONTRIBUTED TO THE EVENT, AN ADDITIONAL REPORT WILL BE SUBMITTED.

Description of Event or Problem · 1

IT WAS REPORTED THAT DOCTOR INDICATED LACK OF PRIMARY STABILITY. NOT OTHER IMPLANT PLACED. TOOTH LOCATION # 13.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1301593 IMP,TSV,3.7,8,MTX,MG DENTAL IMPLANT DZE ZIMMER DENTAL TSVTB8 1231195 00889024019966

Patients

Seq Age Sex Outcome Treatment
1 58 YR Female Required Intervention